20
4. OPERATION PROCEDURE
Mode Key Overview
Using the MODE key.
The MODE key is used to change settings/modes and to monitor settings during operation. Sterilization
temperature, sterilization time, drying temperature and drying time, as well as signal on/off and keypad
tone on/off functions are set/selected with the MODE key
Turn on main power switch.
Sterilizer enters stand-by mode, and
shows current settings.
Press “MODE”.
→
Main display begins flashing
“
”.
(Sterilization temp.setting screen)
Other parameters
・
Turn on main power switch.
・
Press “MODE”.
・
Use
▽△
to toggle thru setting options.
Display Symbol Glossary
Symbol
Meaning
Description
B
BOTTLE
Indicates drain bottle not properly set or installed.
D
DRY TEMPERATURE
Drying temperature setting indicator.
DRY TIME
Drying time setting indicator.
E
END
End of operation indicator.
ERROR
Operational error or malfunction indicator
O
OFF
Signal function off indicator.
ON
Signal function on indicator.
S
STERILIZE TEMPERATURE Sterilization temperature setting indicator.
STERILIZE TIME
Sterilization time setting indicator.
STOP
Operation stop/pause indicator.
SIGNAL
Alert tone setting indicator
Содержание SM201
Страница 2: ......
Страница 17: ...14 3 COMPONENT NAMES AND FUNCTIONS Main Unit Overview 1 ...
Страница 18: ...15 3 COMPONENT NAMES AND FUNCTIONS Main Unit Overview 2 ...
Страница 19: ...16 3 COMPONENT NAMES AND FUNCTIONS Main Unit Overview 3 ...