25
Program “MODE”
Setting Sterilization Temperature and Time
Setting sterilization temp/time
◆
Used to set the sterilization temp/time.
◆
Simply push the ”MODE” key to cancel the set up at any point.
◆
The sterilizer automatically sets in stand-by mode if it is untouched for more than a minute during
set up.
Stand-
by
Turn the breaker on.
The main display blinks sterilization temperature, and the sub
display blinks sterilization time.
Push the “MODE” key.
The main display blinks
, and the sub display flashes
sterilization temperature.
Push the “ENTER” key. At this time, you can change the
sterilization temperature.
Input the sterilization temperature desired by pushing the
▽△
keys.
▽
key
…
To lower temperature.
△
key
…
To raise temperature.
S.
Temp.
Push the “ENTER” key.
The changed sterilization temperature is fixed, and the panel
shows the next function, sterilization time set-up.
Push the “ENTER” key again.
You can now change the sterilization time.
Input sterilization time you desire by
▽△
key.
▽
key
…
To reduce time.
△
key
…
To increase time.
S.
Time
Push the “ENTER” key.
The changed sterilization time is fixed, and the panel blinks the
next function, dry temperature set-up.
D.
Temp.
Push the “MODE” if the dry mode will not be in use. If the dry
mode will be in use, press the “ENTER” key and continue to the
next page.