17
Installation
Initial Set Up
Pour water into the chamber
Pour water into the chamber according to the
gauge on the bottom plate.
Failing to pour adequate water will result in a
interruption of operation. Confirm the water
level after each sample run.
Requirement
SM200
1900
〜
2000ml
SM300
2800
〜
3000ml
SM310
2800
〜
3000ml
See table on right for water volume
information. Using over the maximum level
of water will result in hot water spouting
from the bottle.
SM510
2800
〜
3000ml
Use distilled water for sterilization
Use distilled water to prevent collection of mineral deposits and corrosion of chamber
components. Under ground water is not recommended.
Placing materials into sterilizer
Place the material into the basket and insert
into the chamber. Never block the steam
exhaust port. Always use a basket for
material being sterilized.
High-level
Exhaust Port
Low-level
Содержание SM200
Страница 26: ...24 Program MODE Flowchart ...
Страница 40: ...38 Wiring Diagram SM200 X SM300 Inside of dashed line is option ...