21
Sample Run Examples
How to operate the “STERILIZE” process
Sterilize operation using a waste bag.
①
Be sure to open the mouth of the bag to let steam.
②
Keep the bag approximately 2/3” height of the total depth of the chamber (Example: SM200 is
17.5” deep, so the bag should be placed 11.5” from the bottom). If the bag is placed higher than
this point, steam will not disperse easily and may result in incomplete sterilization, in addition,
the bag can easily cover the exhaust in the upper part of the container.
③
Set the temperature lower than the heat resisting point of the bag.
④
Set time depends on quality and quantity of the material being sterilized. (See chart below)
Example:
Sample Temperature Time
Notes
Gauze
121
℃
30min.
5 dry gauze’s
Petri Dish
121
℃
40min.
30 petri dishes with lid
◆
The above data is only for reference use. The outcome of sterilization varies according to quality
and quantity of substance and types of containers. The result of sterilization shall be checked
using a chemical indicator.
NOTE: Never operate this unit without a basket.
Содержание SM200
Страница 26: ...24 Program MODE Flowchart ...
Страница 40: ...38 Wiring Diagram SM200 X SM300 Inside of dashed line is option ...