31
9. MAINTENANCE
Daily Inspection
Conduct the daily inspection at least once a day according to the following table.
Daily Inspection Table
No.
Place
Item
Criteria
Result
Remarks
1
Cooling fan
Noise
Smooth rotation with no abnormal
noise.
Vibration
No abnormal vibration.
Odor
No offensive odor.
2
Chamber door Opening/
closing
Smooth movement.
Tightness
No gap between the chamber and
the door.
3
Supply gases
Pressure
Process gas
=
0.15 to 0.2 MPa
Purge gas
=
0.2 to 0.3 MPa
4
Gas
connect
port
Gas leak
No gas leak.
5
Vacuum
exhaust
connect port
Tightness
Not loose.
6
Commercial
in-put line
Connection
No loose connectors.
No abnormal temperature rise of
connectors.
7
Vacuum pump Oil level
Within the level lines.
Oil color
Not contaminated.
Noise
No abnormal noise.
Vibration
No abnormal vibration.
Oil leak
No leak.
Connection
wires
No loose connections.
● Prior to the inspections and maintenance, turn off the power switch on the distribution panel for safety.
● Wait until the machine has returned to the normal temperature before maintenance.
● Never disassemble the machine.