39
Operation Method
Program Creation Method
1
Program creation; Step 1 (S01)
Fan
►
10
S01
Rep. start
►
No
S01
⑥
Press the PROGRAM key. The fan rotation speed
selection screen is displayed.
⑦
Press the ENTER key. The fan rotation speed edition
screen is displayed. Set "10" using the "
▼▲
" and
press the ENTER key to decide it. (The next selection
screen can be displayed by pressing the ENTER key
when the setting value is indicated in the selection
screen.)
⑧
Press the PROGRAM key. The repeat initiating step
selection screen is displayed.
⑨
Press the ENTER key. The repeat initiating step edition
screen is displayed. Set "No" using the "
▼▲
" and
press the ENTER key to decide it.
2
Program creation; Step 2 (S02)
TROUBLE
ENTER
CANCEL
DOOR
Set TEMP
S02
S02(un-used)
Set TEMP
►
0
℃
S02
Set time
►
0min
S02
Fan
►
10
S02
①
Keep pressing the PROGRAM key until "SteP2" (S02) is
displayed.
Set TEMP
►
0
℃
S02(un-used)
(Displayed only when nothing is input in Step 2)
②
Press the ENTER key. The setting temperature
selection screen is displayed.
③
Press the ENTER key. The setting temperature edition
screen is displayed. Set "100
℃
" using the "
▼▲
" and
press the ENTER key to decide it.
④
Press the PROGRAM key. The setting period selection
screen is displayed.
⑤
Press the ENTER key. The setting period edition
screen is displayed. Set "30min" using the "
▼▲
" and
press the ENTER key to decide it.
⑥
Press the PROGRAM key. The fan rotation speed
selection screen is displayed.
⑦
Press the ENTER key. The fan rotation speed edition
screen is displayed. Set "5" using the "
▼▲
" and press
the ENTER key to decide it.
Содержание DNF 400
Страница 2: ......
Страница 17: ...13 Operation Method Key Operation Chart of Mode Setting and Program Registering ...