26
Operation Method
Input Program
2
Edit step
The step number currently edited is displayed on the lower column of Sub Display. The details for all
registered steps in use are displayed followed by all un-used steps.
The "Set TEMP" for the next step follows the "Program End". The screen, in this way, displays the
details of respective steps in sequence. The unused steps are displayed at the end if other steps are
used. The display "un-used" is added at the end of step number displayed in the lower column of
Sub Display for un-used steps. All steps subsequent to the step with "un-used" are un-used steps.
•
The setting items are not displayed on the screen depending on the setting conditions shown
below.
It is impossible to change the content of program currently operated. Checking it is possible.
No. Item
Sub
Display
Notes
7
Step deletion
Delete step
S01
Delete the step currently referred to. The
sequence number of each step hereafter
decreases by one.
8
Program end
Program End
S01
Complete program registration/edition.
Setting item
Not displayed in the following condition
Setting temperature
Not displayed when the setting time is set to "End".
Setting time
Always displayed.
Rotation speed of fan
Always displayed.
Repeat initiating step
Not displayed when the period is set to "End" or when the
other step is inserted in the specified repeat interval.
Repeat count
Not displayed when the repeat initiating step is not
displayed, or "No" is set.
Step insertion
Not displayed when steps are left.
Содержание DNF 400
Страница 2: ......
Страница 17: ...13 Operation Method Key Operation Chart of Mode Setting and Program Registering ...