19
Operation Method
Auto Stop Operation
Start operation
AUTO START
FIXED TEMP
AUTO STOP
PROGRAM
STANDBY
END
HEATER
FAN
FUNCTION
PROGRAM
MENU
ENTER
CANCEL
SUBMENU
POWER
TROUBLE
℃
Set TEMP
100
℃↑
Stop in
30min
DOOR
①
Press the ENTER key to decide the setting and
the auto stop operation starts. The blinking
AUTO STOP lamp lights on and the Sub Display
displays the setting temperature and residual time
to operation stop.
②
The countdown of timer is suspended when the
temperature in furnace is 3
℃
or more lower than
the setting temperature, or 6
℃
or more higher
than it. In this case, the Sub Display displays
"Wait" with blinking. The time display on the
right side of "Wait" shows the total waiting time
in operation.
Set TEMP
100
℃
↑
Wait 1min
(Waiting screen)
4
AUTO START
FIXED TEMP
AUTO STOP
PROGRAM
STANDBY
END
HEATER
FAN
FUNCTION
PROGRAM
MENU
ENTER
CANCEL
SUBMENU
POWER
TROUBLE
℃
Stop time
2002/01/01 12:30
DOOR
③
The operation stops when the residual time
counts zero. The END lamp blinks and the Sub
Display displays the operation finish time when
the operation stops.
The wait function is not activated when the auto stop
operation is carried out with "Clock" mode. The
operation stops at specified time.
④
The fan stops at the same time the operation
finishes when "OFF" is selected at fan function.
The setting of fan function ["ON"/"OFF"] takes
priority, regardless of setting for fan operation in the
standby state.
⑤
Press the POWER key to quit operation.