25
Operation Method
Input Program
1
Select program function
•
Check that the power in turned on.
PROGRAM
AUTO START
FIXED TEMP
AUTO STOP
PROGRAM
STANDBY
END
HEATER
FAN
FUNCTION
PROGRAM
MENU
ENTER
CANCEL
SUBMENU
POWER
TROUBLE
℃
Step
S01
Start in
30min
DOOR
①
Press the PROGRAM key. The program menu
starts and step number selection screen is
displayed.
Press the "
▼▲
". The registered step numbers
and the smallest number of un-used step are
displayed in sequence. Select the step number
among them.
The "S01" is displayed when no program is
registered. In this case, the "
▼▲
" are invalid.
②
Press the ENTER key. The selected step
number is decided and the setting item selection
screen is displayed. The "Set TEMP" is
displayed first.
③
Press the CANCEL key to cancel the program
menu.
2
Edit step
•
After the step number is decided, the setting item selection screen is displayed. Select the items
on the Sub Display using the PROGRAM key.
No. Item
Sub
Display
Notes
1
Setting
temperature
Set TEMP
100
℃
S01
Settable value: 0 to 260
℃
(270
℃
maximum)
Use with the temperature of 260
℃
or less
regardless of the setting temperature range.
2
Setting time
Set time
0min
S01
Settable value: 0 min to 59min, 1h00min to
999h59min, or "End"
3
Rotation speed
of fan
Fan
10
S01
Settable value: 1 to 10, OFF
4
Repeat
initiating step
Rep. start
S01
S01
Input the step number of repeat initiating step
or "No".
5
Repeat count
Rep. count
5
S01
Settable value: 1 to 9999, "Endless"
6
Step insertion
Insert step
S01
Add a new step at the position of step
currently referred to. The sequence number
of each step hereafter increases by one.
Содержание DNF 400
Страница 2: ......
Страница 17: ...13 Operation Method Key Operation Chart of Mode Setting and Program Registering ...