pH/ION 3310
Ion concentration
ba77052d01
03/2014
41
Continuing with two-
point calibration
7.
Thoroughly rinse the electrode with distilled water.
8.
Immerse the electrode in standard solution 1.
9.
When calibrating without temperature sensor:
Measure the temperature of the standard solution using a thermom-
eter.
Use
<F2>
/[
⇅
] to select the setting of the temperature.
Use
<
> <
>
to set the temperature.
Use
<F2>
/[
⇅
] to select the setting of the concentration.
10.
Set the concentration of the standard solution with
<
> <
>
and press
<ENTER>
.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
11.
Wait for the end of the AutoRead measurement or accept the calibration
value with
<ENTER>
.
The calibration display for the next standard solution appears.
ISE
0.0
24.8 °C
15.02.2014
12:12
0.01 mg/l
mV
AR
ISE
0.0
24.8 °C
Ag
0.020
mg/l
2
15.02.2014
12:12
Standard
mV
12.
Thoroughly rinse the electrode with distilled water.
13.
Immerse the electrode in standard solution 2.