pH value
pH/ION 3310
26
ba77052d01
03/2014
5.2.4
Manual calibration (ConCal)
Make sure that in the sensor menu,
Buffer
menu, the
ConCal
buffer set is
Use any one to five buffer solutions in ascending or descending order.
21. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
22. If necessary, use
<M>
to finish calibration or
Continue calibrating using the next buffer with
<ENTER>
.
Calibration is automatically completed after the last buffer of a buf-
fer set has been measured. Then the calibration record is dis-
played.
The calibration line is determined by linear regression.
If single-point calibration was set in the menu, the calibration proce-
dure is automatically finished with the measurement of buffer solu-
tion 1 and the calibration record is displayed.
1.
Connect the pH electrode to the meter.
The pH measuring window is displayed.
2.
When measuring without temperature sensor:
– Temper the buffer solution, or measure the current temperature.
– Enter the temperature value with
<
> <
>
.
3.
In the measured value display, select the measured parameter pH or mV
with
<M>
.
4.
Start the calibration with
<CAL>
.
The calibration display appears (voltage display).
pH
180.0
24.8 °C
AutoCal TEC
4.010
15.02.2014
14:15
AR
Buffer
mV
Содержание WTW pH/ION 3310
Страница 76: ...Index pH ION 3310 76 ba77052d01 03 2014...
Страница 77: ......