pH value
pH/ION 3310
32
ba77052d01
03/2014
Calibration record
(example)
5.2.7
Continuous measurement control (CMC function)
The Continuous Measurement Control (CMC function) facilitates to evaluate
the current measured value instantly and definitely.
After each successful calibration the scale of the pH measuring range is
displayed in the measured value display. Here you can very clearly see
whether or not the current measured value is in the calibrated part of the
measuring range.
pH/ION 3310
Ser. no. 11292113
CALIBRATIONpH
15.02.2014 15:55
AutoCal TEC
Buffer 1
4.01
Buffer 2
7.00
Buffer 3
10.01
Voltage 1
184.0 mV
Voltage 2
3.0 mV
Voltage 3
-177.0 mV
Temperature 1
24.0 °C
Temperature 2
24.0 °C
Temperature 3
24.0 °C
Slope
-60.2 mV/pH
Asymmetry
4.0 mV
Sensor
+++
etc...