pH/Cond 3320
Settings
ba77054d01
02/2014
55
The sensor settings are reset under the
Reset
menu item in the menu for cali-
bration and measurement settings. To open it in the measured value display,
press the
<ENTER>
(or
<F1>
/
[Menu]
)
key.
Cond
The following settings for conductivity measurements are reset to the default
settings with the
Reset
function:
The sensor settings are reset under the
Reset
menu item in the menu for cali-
bration and measurement settings. To open it in the measured value display,
press the
<ENTER>
(or
<F1>
/
[Menu]
)
key.
8.3.2
Resetting the system settings
The following system settings can be reset to the default condition:
Slope
-59.2 mV
Man. temperature
25 °C
One point calibration
No
Setting
Default settings
Setting
Default settings
Kal.-Intervall
150 d
Measured parameter
ϰ
Cell constant (C)
(calibrated)
0.475 cm
-1
or 0.840 cm
-1
(nominal cell constant of the conduc-
tivity measuring cell last calibrated)
Cell constant (C)
(set)
0.475 cm
-1
Temperature compensation
nLF
Reference temperature
25 °C
Temperature coefficient (TC) of the
linear temperature compensation
2.000 %/K
TDS factor
1.00
Setting
Default settings
Language
English
Beep
On
Baud rate
4800 Baud
Output format
ASCII
Decimal separator
Dot (xx.x)
Contrast
50 %