Settings
pH/Cond 3320
50
ba77054d01
02/2014
Measurement
/
Measuring cell
/
Type
Measuring cell
used
Cal
Measuring cells whose cell constant
is determined by calibration in the
KCL control standard solution.
Calibration ranges:
0.450 ... 0.500 cm
-1
and
0.800 ... 0.880 cm
-1
The currently valid cell constant is
displayed in the status line.
LR325/01
LR 325/01 measuring cell,
nominal cell constant
0.100 cm
-1
.
The cell constant can be adjusted in
the range 0.090 ... 0.110 cm
-1
.
LR325/001
LR 325/001 measuring cell,
nominal cell constant
0.010 cm
-1
.
The cell constant is permanently set.
man
Any measuring cells with freely
adjustable cell constants in the range
0.250 ... 25.000 cm
-1
.
Measurement
/
Measuring cell
/
Cell const. man
0.250 ...
0.475
25.000 cm
-1
Display and setting option of the cell
constant of any measuring cells
(man).
Measurement
/
Measuring cell
/
Cell const. LR325/
01
0.090 ...
0.100
...
0.110 cm
-1
Display and setting option of the cell
constant of the LR 325/01 measuring
cell.
Measurement
/
Temp. comp. (TC)
/
Method
nLF
Lin
Off
Procedure for temperature compen-
sation (see section 8.1.7 T
This setting is only available for the
measured parameters,
ϰ
and
ρ
.
Measurement
/
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000
...
10.000 %/K
Coefficient of the linear temperature
compensation.
This setting is only available when the
linear temperature compensation is
set.
Menu item
Possible
setting
Description
Содержание WTW pH/Cond 3320
Страница 80: ...Index pH Cond 3320 80 ba77054d01 02 2014...
Страница 81: ......