93
Dissolved Oxygen Transmitter
GB Oxix Manual 1209
FW: 841013-009, 846001-002, 841513-009
D. Upgrade Converter Firmware
1. Select "Firmware" in the menu bar and click "Converter".
2. Locate the "MJK Field FW" directory on the CD-ROM, select the "842xxx-
xxx.hex" fi le (or similar) and click "Open".
3. Select the required
Oxix
®
(option 1, 2, 3 or 4) and click OK when the
"Converter Firmware" dialogue box appears and prompts for selecting the
converter to program.
The converter fi rmware is then transferred, and the converter Flash pro-
gramming is initiated. Expect this process to take a few minutes.
4. When the Flash programming is done, click OK on the PC and the
Oxix
®
Display Unit to return to the
Oxix
®
display.
5. If required you may adjust the
Oxix
®
Display Unit screen contrast by
pressing the two outmost keys simultaneously, press the up/down keys
as required, and save the new setting by again pressing the two outmost
keys simultaneously.