Dissolved Oxygen Transmitter

GB Oxix Manual 1209

FW: 841013-009, 846001-002, 841513-009

D.   Upgrade Converter Firmware

1.  Select "Firmware" in the menu bar and click "Converter".
2.  Locate the "MJK Field FW" directory on the CD-ROM, select the "842xxx-

xxx.hex" fi le (or similar) and click "Open". 

3.  Select the required 

Oxix

® 

(option 1, 2, 3 or 4) and click OK when the 

"Converter Firmware" dialogue box appears and prompts for selecting the 
converter to program.

  The converter fi rmware is then transferred, and the converter Flash pro-

gramming is initiated. Expect this process to take a few minutes.

4.  When the Flash programming is done, click OK on the PC and the 

Oxix

® 

Display Unit to return to the 

Oxix

® 

 display.

5.  If required you may adjust the 

Oxix

® 

Display Unit screen contrast by 

pressing the two outmost keys simultaneously, press the up/down keys 
as required, and save the new setting by again pressing the two outmost 
keys simultaneously.

Содержание MJK Oxix

Страница 1: ...1 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter Installation and User Manual Dissolved Oxygen Transmitter...

Страница 2: ...MC stipulated in Council directive no 89 336 EEC of 3rd may 1989 alteret at directive no 92 31 EEC on the approximation of the laws of the member states relating to electromagnetic compat ibility CE a...

Страница 3: ...ical Sensor Installation 20 Preparation 20 Positioning the Sensor 20 5 Sensor Calibration 21 Zero point Calibration Check 21 2 point Calibration 21 6 Converter 23 Electrical Mounting 23 Power Supply 2...

Страница 4: ...Screen 42 Datalogger 44 Graph Display 46 Password 47 Sensor Name 49 Converter Setup 51 Averaging 52 Units 52 mA Output 53 Cleaning 54 High alarm 56 Low alarm 57 Sensor error 58 Status 59 Factory sett...

Страница 5: ...3 E Install Add Languages 94 Appendix C Front Panel Cut out Drawing 95 Appendix D Log Files 97 Appendix E Input Output Connections 101 Appendix F Quick Guide Oxix Calibration 102 Zero Point Calibratio...

Страница 6: ...6 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter This page intentionally left blank...

Страница 7: ...ter is easy to install and put into service However read this manual first to get familiar with the Oxix transmitter and all its features The equipment must be handled and operated as instructed by th...

Страница 8: ...positioning of the transmitter in liquids distances to walls and other solid faces and installation in tubes 5 Sensor Calibration The Oxix sensor is factory calibrated and requires no re calibration 6...

Страница 9: ...A 1 1 scale drawing of the front panel outline and cut out area for installa tion and mounting purposes Appendix D Log Files An example log file illustrates the format and explains the entries Appendi...

Страница 10: ...orbs the light energy by changing the outer electron s ener gy level The unstable electron then collapses and returns back to its original energy state and at the same time emits the surplus energy as...

Страница 11: ...lay Unit can control up to four MagFlux flow meters SuSix trans mitters and or Oxix transmitters for greater economy space savings and an improved overview of the multiple measurement values Simple to...

Страница 12: ...hes etc User definable Text Up to five lines of text and readings can be configured by the user The graphic display is automatically adjusted to show the largest characters pos sible Alarms can be dis...

Страница 13: ...shock from the Mains sup ply terminals N and L Be careful not to touch these terminals when the Oxix front cover is removed WARNING Incorrect installation or use can lead to body injury and or materia...

Страница 14: ...14 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter This page intentionally left blank...

Страница 15: ...or 24 V AC 50 60 Hz 10 or 115 V AC 50 60 Hz 10 or 230 V AC 50 60 Hz 10 Power consumption Max 10 W Cabinet material Polycarbonate glass reinforced Enclosure rating IP 67 NEMA 6 Temperature range 20 60...

Страница 16: ...3 3 4 Bar Air supply 40 60 psi 2 7 4 Bar Cable 4 x 0 326 mm2 shielded outer diameter 5 0 mm 22 AWG Cable length 10 m 33 ft can be extended up to 600 mm Response time t90 less than 1 second t95 less t...

Страница 17: ...206306 Oxix Converter w display 24 V AC 206307 Oxix Converter without display 10 30 V DC 206308 Oxix Converter without display 230 115 V AC 206309 Oxix Converter without display 24 V AC Oxix Sensor 2...

Страница 18: ...18 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter This page intentionally left blank...

Страница 19: ...xample pumps and other equipment must be avoided The system should be installed to facilitate access to the Oxix transmitter operator panel and sensor without risking personal injury during normal ope...

Страница 20: ...submerged to a fixed depth from the surface In tanks or basins with no or little fluctuation in depth the sensor may be mounted in a fixed position approximately 1 4 below the surface The optimal pos...

Страница 21: ...2 SO3 in a bucket or open container containing 4 liters of clean tap water 1 US gallon 0 9 UK gallon 2 Stir for about 1 minute 3 Lower the sensor into the solution and let it rest at least half an hou...

Страница 22: ...22 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter This page intentionally left blank...

Страница 23: ...dis play unit to gain access to the terminals Mains supply mA D02 D01 DI Display A B C2 N02 C NO N01 PE N L C1 DC GND A B Red Black White Green Shield 521275 Line 2 wire shielded cable 4 wire shielde...

Страница 24: ...al Fuse Ratings Voltage Rating Order no Dimension 230 V AC 0 063 mA T 550030 5 x 20 mm 115 V AC 0 125 mA T 550035 5 x 20 mm 24 V AC 0 5 mAT 550049 5 x 20 mm 10 30 V DC 1 0 AT 550051 5 x 20 mm The tech...

Страница 25: ...age 23 2 Note down the color and position of the wires in the terminal blocks and then loosen the terminal screws 3 Unscrew the two screws that hold the metal cover and then remove it 4 Unscrew the fo...

Страница 26: ...for a 115 V AC configuration 115 V AC dual jumper positions 8 Insert and solder 2 jumpers wires in the positions indicated by the black arrows 9 Turn around the PCB and replace the 63 mAT fuse on the...

Страница 27: ...tails on page 53 Digital Output Oxix has two digital outputs DO1 with an opto light triggered relay and DO2 with an electro mechanical relay They can both be programmed for the following functions Sen...

Страница 28: ...ted with a voltage higher than 10 V DC and de activated with a voltage lower than 5 V DC Digital Input Terminal Designation DI Com DI Max 30 V DC The digital input DI can be programmed for the followi...

Страница 29: ...ire shielded cable display unit display unit Display unit Converter 4 Converter 1 Max four MJK converters with Modbus can be connected to one display unit 521274 2 5 Main supply Supply Fuse 10 30 V DC...

Страница 30: ...D01 DI Display A B C2 N02 C NO N01 PE N L C1 Line 2 wire shielded cable 4 wire shielded cable display unit display unit Display unit Converter 4 Converter 1 Max four MJK converters with Modbus can be...

Страница 31: ...on protocols with other MJK prod ucts like the MJK Magflux flow meter and the MJK Connect data transmitter and can also share display units This section illustrates two configuration examples and incl...

Страница 32: ...n Magflux flow sensor converter with a blind lid connected to a Oxix transmitter with display unit Part numbers for this configuration 2x7xxx MagFlux Flow Sensor 2x7791x MagFlux Connverter with blind...

Страница 33: ...d 2 to slot Display B 3 Connect the twisted shield to slot Display Magflux with Blind Lid 2 wire shielded twisted cable 1 Connect lead 1 to slot Display A 2 Connect lead 2 to slot Display B 3 Connect...

Страница 34: ...lind lid connected to a Mag flux flow sensor with a blind lid and common remote Display Unit Part numbers for this configuration 206510 Oxix Sensor 206304 6 SuSix Oxix Converter 2x7xxx MagFlux Flow Se...

Страница 35: ...wisted shield to slot Display 4 wire shielded twisted cable 1 Connect lead 1 to slot Display 2 Connect lead 2 to slot Display A 3 Connect lead 3 to slot Display B 4 Connect the twisted shield to slot...

Страница 36: ...36 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter This page intentionally left blank...

Страница 37: ...ightened 4 All cable glands are tightly secured Initial Power up 1 Turn on power 2 Within 60 seconds the Oxix sensor will be found by the automatic sen sor detection program During this process you mi...

Страница 38: ...p configuration and normal opera tion Display and keyboard for reading and programming The display shown above will in the remaining part of this manual be stylized to appear like this Specify main Sc...

Страница 39: ...Set log interval 0020 sec Back Select Datalogger Select line Linie 1 Sensor name Linie 2 Oxygen sat Linie 3 Linie 4 Linie 5 Back Select OK Specify main Screen Select line content Not in use Sensor na...

Страница 40: ...ing all four keys simultaneously It is a so called coldboot where all displays are refreshed and all settings configurations are kept with only one exception the language setting This feature is espec...

Страница 41: ...t Linie 3 Linie 4 Linie 5 Back Select OK Specify main Screen Select line content Not in use Sensor name Measurement Temperature Clock Back Select OK Type Back Oxix1 Oxygen SAT Select Set up Main Menu...

Страница 42: ...en sat Linie 3 Linie 4 Linie 5 Back Select OK Specify main Screen Select line content Not in use Sensor name Measurement Temperature Clock Back Select OK Type password 0000 Back Select Activ Deac Back...

Страница 43: ...ure The temperature may be displayed in degrees Celcius C or in degrees Fahrenheit F See also page 52 Clock Time and date The size of the display lines will automatically increase or decrease as the n...

Страница 44: ...ensor name Measurement Temperature Clock Back Select OK T B Oxix1 Oxygen SAT Select Set up Main Menu Converter Overview Menu Oxix1 91 9 Susix1 578 mg L Back Select Display Display Overview Menu Displa...

Страница 45: ...ame memory of 20 000 entry points The log data can be shown on the Display Unit or be stored in an external CSV file MJK Field Link software is needed for transfer of data into CSV file format via the...

Страница 46: ...he Display Unit then shows the Graph screen To return to the Main Menu display screen select esc by pressing the up down keys simultaneously 100 2008 09 21 11 25 43 84 Oxix1 Oxygen 98 Select Graph The...

Страница 47: ...lect OK Type p Back Oxix1 Oxygen SAT Select Set up Main Menu Converter Overview Menu Oxix1 91 9 Susix1 578 mg L Back Select Display Display Overview Menu Display setup Sensor name Aeration Tank Inlet...

Страница 48: ...protection Selecting Activate means that a password must be en tered to change vital settings Deactivate disables password protection If your current password is lost or forgotten the password protec...

Страница 49: ...sor name Converter setup Display setup Back Select OK Set log Back Select line Linie 1 Sensor name Linie 2 Oxygen sat Linie 3 Linie 4 Linie 5 Back Select OK Specify main Screen Select line content Not...

Страница 50: ...Main menu 3 Press the up down keys to change the highlighted character and then press to proceed to the next character 4 Continue with until all numeric and alphabetical characters have been set 5 Pr...

Страница 51: ...2 Oxygen sat Linie 3 Linie 4 Linie 5 Back Select OK Specify main Screen Select line content Not in use Sensor name Measurement Temperature Clock Back Select OK T B Oxix1 Oxygen SAT Select Set up Main...

Страница 52: ...the menu is not available or not applicable Enab Yes No Back Averaging 0001 sec from 1 sec to 120 min Back Select Temperature unit C F Back Select OK Temperature This menu is for display only Units Th...

Страница 53: ...tput 4 mA 00 00 mg l Back Select OK Yes Enable mA output Yes No Back Select OK c Temperature unit C F Back Select OK Temperature This menu is for display only Select cleaning interval 060 Min 20 255 M...

Страница 54: ...ck Select OK OK only Clea Yes No Back Select cleaning interval 060 Min 20 255 Min Back Select Cleaning Interval Use DI to clean Clean now Maintenance Back Select OK Interval Off On Back Select OK Use...

Страница 55: ...seconds The measurement delay recovery time may also be used to coordinate with other processes The delay merely freezes the last reading before cleaning begins A typical value would be 60 seconds Cl...

Страница 56: ...o Back Delay Back Electro On Setpoint 7 00 Back Select High Alarm Relay function NO NC Back Select Off Setpoint 6 75 Back Select Electro mechanical relay Opto relay Display only Not in use Back Select...

Страница 57: ...elay fun NO NC Back Low Alarm Electro mechanical relay Opto relay Display only Not in use Back Select OK On Setpoint 5 50 Back Select Relay function NO NC Back Select Off Setpoint 5 75 Back Select Del...

Страница 58: ...ver the 4 20 mA range has been exceeded Sensor error Electro mechanical relay Opto relay Display only Not in use Back Select OK Electro mechanical relay Opto relay Display only Delay 00100 ms Back Sel...

Страница 59: ...Back Select OK Status DO 1 x Selected function DO 1 Selected function DI Selected function mA output 10 234 mA Back Select OK Pending warnings xxx xxx Back Select OK Pending information xxx xxx Back...

Страница 60: ...d check box DI Selected function The selected function linked to digital input DI de activated an empty check box A selected function could for example be Cleaning mA output 10 234 mA The present anal...

Страница 61: ...Back Select Service menu Sensor Data Sensor Calibration Converter SW Version Product data Reset Counter Time Internal meas cal Calibrate mA Test Primary value Read eventlog Comunication log Back On S...

Страница 62: ...rictions but certain parameters can only by changed after a password has been typed Sensor Data Sensor Calibration Converter SW Version Product data Reset Counter Time Internal meas cal Calibrate mA T...

Страница 63: ...on 1 point calibration 2 point calibration Cancel Back Select OK OK Clea No Yes Back Confirm No Yes Back 1 point 2 point Set actua Back Type password 00000 Back Select Service menu Sensor calibration...

Страница 64: ...Bootload 840020 002 684 Back Product data Product ID 000000000000 Back Select HW type no 0510305004 Back Select Track no 00000 Back Select Product type no 00000206305 Back Select et actual measuremen...

Страница 65: ...k Inter View Vcc Back Calibrate Back Reset counter time Reset Counter 1 System Runtime 2304 Since Reset 2304 Runtime days 0 Back Select OK et actual measurement 10 20 mg l k Select OK Reset Counter Ti...

Страница 66: ...e personnel use only The supply voltage Vcc and the reference voltage Vref can be viewed and adjusted Calibrate Vcc Calibrate V_ref Back Select OK Internal meas cal View Vcc Back Select OK Calibrate V...

Страница 67: ...value mg l Back Select Test primary value Enable test Yes No Back Select Current Set fixed current Calib current DAC zero current DAC top current Back Select OK Calib current uA Back Select OK Calibr...

Страница 68: ...mA 4 mA 12 mA 20 mA Maximum mA Back Select OK DAC zero current uA Back Select OK DAC top current Back Select OK No Hour Event Extension 1 154 0 0 2 154 0 0 3 154 0 0 4 154 0 0 Back Select Read eventlo...

Страница 69: ...154 0 0 7 154 0 0 8 154 0 0 Back Select No Hour Event Extension 13 154 0 0 14 154 0 0 15 154 0 0 16 154 0 0 Back Select No Hour Event Extension 9 154 0 0 10 154 0 0 11 154 0 0 12 154 0 0 Back Select N...

Страница 70: ...es Back Select Ok Modbus Rx ok 4950 Rx CRC error 4564 Rx paritet error 4545 Rx adr error 4566 Rx wrong id 4950 Tx count 3454 Back Select OK Communication log Sensor Rx ok 4950 Rx CRC error 4564 Rx par...

Страница 71: ...54 OK log 4950 4564 4545 4950 3454 OK Autodetect sensor type Use Not in use Back Select OK Mains supply frequency 50 Hz 60 Hz Back Select OK Autodetect sensor type Mains supply frequency Mains supply...

Страница 72: ...ect OK Specify main Screen Select line content Not in use Sensor name Measurement Temperature Clock Back Select OK Type pa Back Oxix1 Oxygen SAT Select Set up Main Menu Converter Overview Menu Oxix1 9...

Страница 73: ...k Select OK Language GB English US English Dansk Deutsch Espa ol Fran ais Italiano Nederlands Norsk Suomalainen Svenska Back Select OK Scan Back Set Clock Language Devic Back Language Set Clock Modbus...

Страница 74: ...Dansk Deutsch Espa ol Fran ais Italiano Nederlands Norsk Suomalainen Svenska Back Select OK Ch Devic Back Scanner Back Set Clock Language Add device Devices found Device 1 Back Select OK Ch Chan Back...

Страница 75: ...Change device 1 Back Select OK Remove device Device 1 Back Select OK Remove device Change Display ID No 1 Back Select OK Change display ID no Number of Displays 1 Back Select OK 10 53 12 ct When sever...

Страница 76: ...vice 1 to Device 2 as follows 1 Press OK and use the up down keys to change 1 to 2 2 Press OK The Oxix converter ID is now changed to 2 and more converters can be added to the Display Unit with the Ad...

Страница 77: ...Displays 1 Back Select OK 1 Press the up down keys to highlight the required option and press OK Important The data logger will be reset and all devices will be deleted from the dis play Unit This is...

Страница 78: ...78 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter This page intentionally left blank...

Страница 79: ...ygen Transmitter GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 10 Mechanical Dimensions Sensor 1 4 tube fitting 1 1 2 NPT 112 mm 4 2 Measuring element 1 4 tube Cable to Oxix Converter 130 mm...

Страница 80: ...80 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter Converter and Display Unit 150 144 162 156 M20 186 89 83 183 85 79...

Страница 81: ...ransmitter GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Immersion Fitting Oxix sensor Wall bracket for extension Oxix Immersion fitting order no 155235 Extension for universal bracket Unive...

Страница 82: ...82 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter This page intentionally left blank...

Страница 83: ...nt is scratched by for example the use of a steel brush during cleaning 3 Dry the sensor with a soft cloth 4 Immerse the Oxix sensor into the liquid to be measured and leave it for at least 15 minutes...

Страница 84: ...84 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter This page intentionally left blank...

Страница 85: ...he low set point has been de tected mA connection Alarm The mA connection is not connect ed If the current out is used deactivate mA Current out to stop this alarm Errors Calibration Error 1 Perform c...

Страница 86: ...have been changed Sensor supply Error The sensor supply is too low or too high Check the sensor cable connection System Error Internal VCC error See the event log for details System error Error The sy...

Страница 87: ...but not finished Flash programming done Programming is finished Flash crc error Cyclic redundancy check error Unknown command flash user text Unvalid command issued Flash address overrun A write error...

Страница 88: ...88 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter This page intentionally left blank...

Страница 89: ...log data must be saved in a file before transferring and installing new firmware See 2 Save Log Data overleaf 2 Note down all display and converter settings before upgrading new firmware 3 The upgrad...

Страница 90: ...B connector on the rear of the front panel 4 Connect the other end of the USB cable max 4 7 metres long to the PC 5 Select File in the menu bar and click Mount USB If the connection is successful a D...

Страница 91: ...Display Firmware Important Always upgrade the display firmware before you upgrade the converter software 1 Select Firmware in the menu bar and click Display 2 Locate the MJK Field FW directory on the...

Страница 92: ...xix Display Unit restarts and displays the MJK logo the new link status is displayed and the Display Unit informs that the PC is once again connected Click OK on the PC and the Oxix Display Unit to re...

Страница 93: ...irmware dialogue box appears and prompts for selecting the converter to program The converter firmware is then transferred and the converter Flash pro gramming is initiated Expect this process to take...

Страница 94: ...ages none follow ing a firmware upgrade You can now either add or delete the required languages 3 Click Add select the required language text file or language pack in the MJK Field Display Unit text f...

Страница 95: ...Transmitter GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Appendix C Front Panel Cut out Drawing The dotted line indicates the front panel outline and measures 155 x 145 mm Cut out area Fro...

Страница 96: ...96 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter This page intentionally left blank...

Страница 97: ...The example log file shown below is the result of a CSV file from the converter see Save Log Data procedure on page 91 having been opened in with Microsoft Excel spread sheet utility The example below...

Страница 98: ...percent 95 mg l milligrams per litre 96 kg l kilograms per litre 97 g l grams per litre 130 ft3 h cubic feet per hour 138 l h litres per hour 139 ppm parts per million 246 NTU Nephelometric Turbidity...

Страница 99: ...imum value for graph Precision The precision of the SI value Column J The sensor name as entered by the operator Time Date and time in Central European Standard Time CEST UTime The time in UNIX format...

Страница 100: ...513 009 Dissolved Oxygen Transmitter To plot the log file s data proceed as follows 1 Start Excel and open the csv log file 2 Sort by Index click the Diagram Guide tool in the toolbar and select the r...

Страница 101: ...nd outputs DI and DO can be interconnected to external equipment to achieve a number of functions such as alarms counter reset flow direction indication etc The two schematics below illustrate a simpl...

Страница 102: ...be done independently of each other It is not usually necessary to calibrate We recommend checking the zero position every 6 to 12 months The calibration is saved in the converter and you will need t...

Страница 103: ...conducted as follows The sensor is set for a zero point solution The current measurement is shown in line 2 0 10mg l in this example Set the desired measuring value on the first line using the up and...

Страница 104: ...ally a reference sensor is used to verify the measurement For non oxygen solutions a variation of 1 to 5 between two sensors can occur even when the sensors are placed right next to each other Air bub...

Страница 105: ...arrow keys The interval is from 0 500 to 1 500 Save the current correction factor setup by pressing OK When measurement is done in the ppm or mg litre units a salinity correction is required not when...

Страница 106: ...Factory Calibration Zero point Calibration 0 00 mg l Oxygen 0 10 mg l Set actual measurement Back Select OK The sensor should be reset before calibrating zero point or span by using Reset Factory Cal...

Страница 107: ...the concentration will be higher towards to the surface If you need a reference sample with a dissolved oxygen concentration close to zero approx 0 02 ppm the reference sample must be made with demine...

Страница 108: ...k D Select line Linie 1 Sensor name Linie 2 Oxygen sat Linie 3 Linie 4 Linie 5 Back Select OK Specify main Screen Select line content Not in use Sensor name Measurement Temperature Clock Back Select O...

Страница 109: ...log interval 0020 sec Back Select Datalogger Sensor name Oxix 1 Back Select Sensor name e ame at OK Screen Conve Averaging Units mA output Cleaning High Alarm Low Alarm Sensor Error Status Factory set...

Страница 110: ...nsmitter Main Menu Overview continued 3 me Converter setup Language Set Clock Modbus Factory settings Display SW version Back Select OK Display setup Averaging Units mA output Cleaning High Alarm Low...

Страница 111: ...111 Dissolved Oxygen Transmitter GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 This page intentionally left blank...

Страница 112: ...g l ppm Back Select OK Averaging Units mA output Cleaning High alarm Low alarm Sensor error Status Factory settings Service menu Back Select OK 4mA output 4 mA 0 Back S When n a is displayed the menu...

Страница 113: ...display only Clean now Yes No Back Cle Select cleaning interval 060 Min 20 255 Min Back Select Cleaning Interval Use DI to clean Clean now Maintenance Back Select OK Interval Off On Back Select OK Use...

Страница 114: ...issolved Oxygen Transmitter Converter Setup Menu Overview continued 3 Clean now Yes No Back Select OK Reset Maintenance Yes No Back Select OK Maintenance Runtime 1000 hour Count 500 Back Select OK Cle...

Страница 115: ...Back Select High Alarm Relay function NO NC Back Select Off Setpoint 6 75 Back Select Electro mechanical relay Opto relay Display only Not in use Back Select OK Delay 00100 ms Back Select Electro mec...

Страница 116: ...Opto relay Display only Not in use Back Select OK On Setpoint 5 50 Back Select Relay function NO NC Back Select Off Setpoint 5 75 Back Select Delay 00100 ms Back Select Opto relay Display only On Set...

Страница 117: ...y Display only Not in use Back Select OK Electro mechanical relay Opto relay Display only Delay 00100 ms Back Select Delay 00100 ms Back Select Delay 00100 ms Back Select Relay function NO NC Back Sel...

Страница 118: ...nu Pending errors xxx xxx Back Select OK Factory setting Recover factory settings Cancel Default Force to SuSix Force to Oxix Back Select OK Status DO 1 x Selected function DO 1 Selected function DI S...

Страница 119: ...119 Dissolved Oxygen Transmitter GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 This page intentionally left blank...

Страница 120: ...entlog Comunication log Back On Stock Autodetect sensor type Main supply frequency Back Select OK Sensor data Sensor serial number 30487 Sensor software version 2 Back OK Set actual measurement 10 20...

Страница 121: ...000 Back Select HW type no 0510305004 Back Select Track no 00000 Back Select Product type no 00000206305 Back Select measurement 10 20 mg l elect OK calibration calibration ation ation elect OK OK Cle...

Страница 122: ...ntinued 3 a 00000000 10305004 00000 00206305 Calibrate Vcc Calibrate V_ref Back Select OK Internal meas cal View Vcc Back Select OK Calibrate Vcc Back Select View V_ref Back Select OK Calibrate Vref B...

Страница 123: ...current DAC top current Back Select OK Calib current uA Back Select OK Calibrate mA Current uA Back Select OK Set fixed current Release mA Minimum mA 4 mA 12 mA 20 mA Maximum mA Back Select OK DAC ze...

Страница 124: ...k Select No Hour Event Extension 9 154 0 0 10 154 0 0 11 154 0 0 12 154 0 0 Back Select No Hour Event Extension 17 154 0 0 18 154 0 0 19 154 0 0 20 154 0 0 Back Select Recover Factory Settings Cancel...

Страница 125: ...ved Oxygen Transmitter GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Service Menu Overview continued 6 nsor type t OK Mains supply frequency 50 Hz 60 Hz Back Select OK nsor type Mains supply...

Страница 126: ...Hour AM PM Back Select OK Language GB English US English Dansk Deutsch Espa ol Fran ais Italiano Nederlands Norsk Suomalainen Svenska Back Select OK Scanner Back Set Clock Language Ad Device 1 Back L...

Страница 127: ...evice addresse Device 1 Back Select OK Add device Change Device addresse Remove Device Change Display ID No Back Select OK Scanner Back ck Modbus Add device Devices found Device 1 Back Select OK Chang...

Страница 128: ...rview continued 3 Cancel Delete Log Restore default Language Restore factory setting Back Select OK Display 841013 009 SW Aug 15 2008 15 14 20 Language 841513 009 Back Factory settings Display SW vers...

Страница 129: ...129 Dissolved Oxygen Transmitter GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009...

Страница 130: ...130 GB Oxix Manual 1209 FW 841013 009 846001 002 841513 009 Dissolved Oxygen Transmitter...

Страница 131: ...rm 56 hotline 7 I Important 19 30 index 8 injury 13 info 86 initial checks 37 input digital 28 installation 19 20 Internal Meas Cal 66 introduction 7 Index A accessories 17 accuracy 15 activate deacti...

Страница 132: ...Device 75 Autodetect sensor type 71 Averaging 52 Back on Stock 70 Calibrate mA 67 Change controller ID 76 Change device address 76 Change password 48 Cleaning 54 Clock 43 Communication log 70 Convert...

Страница 133: ...ter Time 65 response time 16 RG 17 RS 485 15 RTU 12 ruthenium 10 S safety 13 SCADA 12 Sensor data 63 error 58 Name 49 service 83 Service Menu detailed 62 digest 61 Set Clock 74 software 9 89 specifica...

Страница 134: ...red trademarks of MJK Automation A S 2013 Xylem Inc 3 05 ES MagFlux Manual 02 2013 Liability MJK Automation A S are liable to the common rules of Danish law on product liability however the liability...

Отзывы: