33
Dissolved Oxygen Transmitter
GB Oxix Manual 1209
FW: 841013-009, 846001-002, 841513-009
Wiring Schematic #1
for " Magflux
®
with Blind Lid and Oxix
®
with Display Unit" from the opposite
page (page
32
).
Oxix
®
with Display Unit
2-wire, shielded twisted cable
1. Connect lead "1" to slot "Display A"
2. Connect lead "2" to slot "Display B"
3. Connect the twisted shield to slot
"Display -"
Magflux
®
with Blind Lid
2-wire, shielded twisted cable
1. Connect lead "1" to slot "Display A"
2. Connect lead "2" to slot "Display B"
3. Connect the twisted shield to slot
"Display -"
1 2 3 4 5 6 7 8 9
Mains
supply
-mA-
-D02-
-D01-
-DI-
-Display-
+
+
A
B
C2 N02
C NO
N01
PE
N
L
-
-
C1
Line
2 wire
shielded cable.
4 wire
shielded cable.
displa
y unit
displa
y unit
Display unit
Converter #4
Converter #1
Max. four MJK converters with Modbus
can be connected to one display unit:
521274
2,5
Main supply
Supply
Fuse
10-30
V DC
1 A
T
24
V A
C
500 mA
T
115
V A
C
125 mA
T
230
V A
C
63 mA
T
Sensor supply
Fuse 200 mA T
+DC
GND A
B
Red
Black
White
Gr
een
Shield
521275
www.mjk.com
DK: +45 45 56 06 56
NO: +47 69 20 60 70
SE: +46 53 31 77 50
NL: +31 251 672171
USA: +1 847 482 8655
AUS: +61 3 9758 8533
1 2 Shield
1 2 Shield