19
Dissolved Oxygen Transmitter
GB Oxix Manual 1209
FW: 841013-009, 846001-002, 841513-009
4. Sensor Installation
WARNING!
The Oxix
®
transmitter
must not be powered until the transmitter is
correctly mounted and connected!
WARNING!
Connections to the Oxix
®
transmitter
must not be disconnected while
the transmitter is powered.
Important!
To ensure safe and optimal functionality of the Oxix
®
transmitter,
the instructions below must be followed closely.
• The sensor must be mounted in a location where it is possible to perform
a representative dissolved oxygen sat. measurement.
• Fittings and mountings must be capable of carrying the sensor's weight
(0,77 kg / 1.7 lbs), and vibrations from for example pumps and other
equipment must be avoided.
• The system should be installed to facilitate access to the Oxix
®
transmitter
operator panel and sensor without risking personal injury during normal
operation, maintenance, cleaning of sensor, etc.
• The Oxix
®
transmitter must be connected to a mains outlet that complies
with national and local regulations.
• All cables must be installed, connected and mounted to minimize me-
chanical impact and electrical influence from other installations.
• The sensor cable’s maximum tensile load is adjusted to the sensor weight.
• The sensor is a electro-mechanical instrument. Be careful not to damage
the sensor lens. The accuracy will deteriorate if foreign material accumu-
lates on it, or it is subject to abrasive cleaning procedures. Only use a soft
cloth for cleaning. NEVER EVER a steel brush!
• All fuses are located on the Oxix
®
converter’s motherboard.