Section 5: System Errors
5-6 | 14/11/2018
PHET_OP_GB Version 7© Xstrahl Ltd. All Rights Reserved
Error Code Table
No.
Message Displayed
Description
80
Anode Decay too Slow
Generator error. Displayed in bipolar systems
if the decay of the anode kV at the end of the
exposure takes too long. The system status
will remain as X-Ray ON until the kV threshold
is reached.
81
Interlock Dropout
Generator error. Displayed if an exposure is
terminated by the loss of the interlock signal.
Latching error.
82
Interrupted by HS Temp
Generator error. Displayed if an exposure is
terminated by the heatsink thermostat.
Latching error.
83
Not Used
Generator error. Displayed if the
unipolar
or
bipolar
tables are incompatible.
84
Dual Door Interlock Failure
Displayed if the two door sensors fail to
operate together when configured to do so.
This will prevent X-Ray ON.
85
Overriding Mandatory WU
Generator error. Not used in this system.
86
Exposure Param Error
Generator error
87
Anode Residual kV
Generator error. Displayed in bipolar systems
if the anode generator tank kV exceeds a
certain value in X-Ray OFF.
88
Anode Residual mA
Generator error. Displayed in bipolar systems
if the anode generator tank mA exceeds a
certain value in X-Ray OFF.
89
No mA at Switch On
Generator error. Displayed if the mA
measured remains at zero after a specified
time.
91
Shutdown by Residual kV
Generator error. Latching error. Displayed if a
residual kV error has terminated X-Rays.
92
Shutdown by Residual mA
Generator error. Latching error. Displayed if a
residual mA error has terminated X-Rays.
93
Timer Interrupt Late
Generator error
94
Generator Not Ready
Generator error
95
High Energy Discharge
Generator error
107
Generator Interlock
Problem
Displayed if the control of the interlock relay
does not result in the correct response from
the generator.
108
App Factor Required
Displayed when the applicator is selected for a
Dose
exposure and the applicator factor for
the selected applicator is out of range. (Must
be greater than or equal to 0.8 and less than
or equal to 1.2.)
109
Dose Requested too High
Displayed when the dose entered in a
Dose
treatment exceeds the set dose limit. This
error will persist until a compliant dose is
entered.
110
LCD Failure
Displayed if the TP2 fails to read data from the
LCD. This error will persist until rectified.
Содержание RADiant
Страница 1: ...RADiant Therapy System Operator Manual Language English...
Страница 2: ......
Страница 10: ...RADiant Operator Manual About Xstrahl viii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 14: ...RADiant Operator Manual ii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 16: ...RADiant Operator Manual iv 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 36: ...RADiant Operator Manual 4 2 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 45: ......