Section 3: Xstrahl RADiant Therapy System Operation
PHET_OP_GB Version 7 ©2017 Xstrahl Ltd. All rights reserved
14/11/2018| 3-3
5.
Press
X-Rays On
Using Fisica:
1.
Enter the
username and password
at the prompt and select
OK
on the
log on screen. Once logged in, the username will be displayed on the
left hand side of the screen. The warm-up can only be conducted if the
system is synchronised.
2.
Select
Tools
then
Warm-Up
to launch the warm-up screen. The system
will default to the required Warm-Up. The radio buttons can be used to
select a different warm-up (
3.
Enter the operator name
4.
Click
Apply
. Apply checks the filter, applicator and the status of the
system.
5.
Turn the operator pod key switch to
HT
(position 3).
6.
Press X-Rays On
Note:
Do not perform interlock checks (e.g. on doors or the key switch) during a
warm- up.
Note:
Warm-ups will take longer if interrupted
Note:
If a warm up has been completed another cannot be requested without
removing removing power to the system.
3.3
Interrupting an Exposure During the Beam Ramp Up
If the X-Ray beam is interrupted in any way (for example Door opened, E-
stop,
Ray off) during the kV ramp (that is, without the mA to load the kV), it
will take 5-8 seconds for the kV to drop below the 5kV level. In this
situation the HT ON indication continues (even though the beam is
interrupted) until the kV drops below 5kV. This is the correct operation of
the system in this circumstance.
If possible, the beam should only be interrupted when the kV has ramped all
the way up and the mA has started or stabilised. This can be seen in the
‘actual’ kV and mA values on the exposure
screen.
1. Operator name must be at least three alphanumeric characters.
3.4
Power Down Procedure
The
Xstrahl X-Ray Therapy System
can be left powered on with the operator
pod key switch in
standby
(position 2). During the clinical session this
Содержание RADiant
Страница 1: ...RADiant Therapy System Operator Manual Language English...
Страница 2: ......
Страница 10: ...RADiant Operator Manual About Xstrahl viii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 14: ...RADiant Operator Manual ii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 16: ...RADiant Operator Manual iv 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 36: ...RADiant Operator Manual 4 2 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 45: ......