Section 3: Xstrahl RADiant Therapy System Operation
3-2 | 14/11/2018
PHET_OP_GB Version 7© Xstrahl Ltd. All Rights Reserved
3.2
System Warm-Up
The X-Ray tube requires a daily warm-up to be conducted. The system
performs a fully automated warm-up.
Note:
If the warm-up sequence is bypassed, it is recorded in the system log
and will
invalidate the manufacturer’s warranty on the X
-Ray tube
The following table provides the standard warm-up times for the system:
Table 3-1: Warm up times
Xstrahl RADiant Therapy Warm-Up Times
Xstrahl System
Short Warm-Up
Long Warm-Up
80kV
1.15 minute
20.25 minutes
The long warm up is defaulted to if the system is left inactive for >2 days.
Once the system is powered on, a warm-up should be conducted.
To conduct a warm-up:
1.
Fit the warm-up filter and a small aperture applicator onto the system.
The system requires an applicator to be fitted to the sub-tube assembly,
but does not require a specific applicator.
2.
Double click the
icon (either Fisica or Concerto).
Both Xstrahl software applications provide the ability to conduct a system
warm-up. If there is a problem with the communication between the PC
and TP2 hardware, an error dialog will be displayed.
Please follow the Concerto or Fisica heading, depending on which
interface you are using to conduct the warm-up.
Using Concerto:
1.
Log on to Concerto. Select
Treatment
then
Start a Warm-Up
Figure 3-1: Warm-Up Window
The system will default to the warm-up required by the system. Use the
radio buttons to select a different warm-up if required.
2.
Enter the
operator name
into the Operator field
1
3.
Click
Apply
. Apply checks the filter, applicator and the status of the
system.
4.
Turn the operator pod key switch to
HT
(position 3).
Содержание RADiant
Страница 1: ...RADiant Therapy System Operator Manual Language English...
Страница 2: ......
Страница 10: ...RADiant Operator Manual About Xstrahl viii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 14: ...RADiant Operator Manual ii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 16: ...RADiant Operator Manual iv 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 36: ...RADiant Operator Manual 4 2 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 45: ......