Section 2: Xstrahl RADiant System Description
2-10 | 14/11/2018
PHET_OP_GB Version 7© Xstrahl Ltd. All Rights Reserved
2.5
Warning Lights and Door Interlock
The RADiant Therapy System has warning lights on access to the
treatment room.
Controlled Area X-rays illuminated (Yellow) when the system is powered.
Do Not Enter illuminated (Red) and flashing when X-rays are on.
Door interlock switches located on the access door to the treatment room.
Figure 2-11: Warning Lights (at access to treatment room)
2.6
Cleaning and Disinfecting
WARNING
: Always carry out cleaning procedures with the
mains power switched off. When using disinfectants, do not use
agents that when mixed with air produce flammable or
explosive vapours. Do not subject this equipment to liquid
spills, ingress of liquids or harmful substances
Carry out cleaning and disinfecting as necessary. Dust metallic parts as
required. If soiling or more stubborn stains exist, use a non-abrasive
cleaning agent and apply with a damp,
not wet
, cloth.
Use a cleaning wipe such as Clinell Universal Wipes, for surface
disinfection of the medical device (NHS Supply Chain: VJT118), and dry
after use.
Follow the manufacturer’s recommended instructions supplied with
cleaning agent or disinfectant.
Содержание RADiant
Страница 1: ...RADiant Therapy System Operator Manual Language English...
Страница 2: ......
Страница 10: ...RADiant Operator Manual About Xstrahl viii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 14: ...RADiant Operator Manual ii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 16: ...RADiant Operator Manual iv 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 36: ...RADiant Operator Manual 4 2 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 45: ......