Section 5: System Errors
PHET_OP_GB Version 7 ©2017 Xstrahl Ltd. All rights reserved
14/11/2018| 5-5
Error Code Table
No.
Message Displayed
Description
57
Shutdown by Safety2
Displayed if the X-Ray beam is stopped by the
second door/room interlock contact being
opened. Latching error requiring a new X-Ray
ON signal to be reset.
58
Shutdown by Cooler Flow
Displayed if the X-Ray beam is stopped by the
cooler flow contact being opened. Latching
error requiring a new X-Ray ON signal to be
reset.
59
Shutdown by Cooler Temp
Displayed if the X-Ray beam is stopped by the
cooler temp contact being opened. Latching
error requiring a new X-Ray ON signal to be
reset.
60
Contactor Closed at Power
Up
Displayed at start-up if the
safety
contactor is
engaged but not enabled.
64
System has RESET
Generator error. Displayed if the processor in
the generator has reset.
66
Key Turned During
Exposure
Displayed if the X-Ray beam is stopped by the
control pod key being moved from
HT
(position 3), the X-Ray enable position.
Non-latching error.
67
High kV Demand in XOff
Generator error
68
High mA Demand in XOff
Generator error
71
Droop Cal. too High
Generator error
72
DAC Offset too High
Generator error. Displayed if the
digital
to
analog converter
(DAC) offset voltage is too
high. This will inhibit an exposure.
73
DAC Range Problem
Generator error. Displayed if the
digital
to
analog converter
(DAC) range data is
incorrect. This will inhibit an exposure.
74
ADC Zero Offset too High
Generator error. Displayed if the
analog
to
digital converter
(ADC) offset voltage is too
high. This will inhibit an exposure.
75
ADC Range Problem
Generator error. Displayed if the
analog
to
digital converter
(ADC) range data is incorrect.
This will inhibit an exposure.
76
ADC Calibration Lost
Generator error. Displayed if the
analog
to
digital converter
(ADC) calibration is lost. This
will inhibit an exposure
77
Bad Generator Table in
PROM
Generator error. Displayed if the generator
type data is corrupt. This will inhibit an
exposure.
78
kV Breakdown Lockout
Generator error. Displayed if the X-Rays have
been terminated by three events at
successively lower kV values. This will inhibit
an exposure.
79
X-Ray Off | / O Problem
Generator error. Displayed if the generator
receives an X-Ray ON initiation, but the X-Ray
OFF interlock line is not enabled (sourced with
current).
Содержание RADiant
Страница 1: ...RADiant Therapy System Operator Manual Language English...
Страница 2: ......
Страница 10: ...RADiant Operator Manual About Xstrahl viii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 14: ...RADiant Operator Manual ii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 16: ...RADiant Operator Manual iv 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 36: ...RADiant Operator Manual 4 2 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Страница 45: ......