The Future Starts Now

 

 

OXY 70 

4. Dissolved oxygen measurement 

4.1  Before starting 

In order to reduce measurement errors and achieve the greatest possible accuracy, observe the following 

rules before you begin: 

 

The sensor must be calibrated 

 

The sensor must be vertical with the membrane downwards 

 

Remove the protective cap 

 

The sensor must be at the same temperature of the sample to be analysed, if necessary leave the 

probe immersed in the sample until the thermal equilibrium is reached 

 

4.2  Methods of measurement 

This instrument can work in two measurement modes: 

 

Dissolved O

2

 Saturation

 expressed in % 

 

Dissolved O

2

 

Concentration 

expressed in mg/L, equivalent of ppm 

mg/L = ppm

 

During measurement press 

  key to change measuring method. 

 

4.3  Making the measurement 

Remove  the  electrode  from  the  protective  cap,  rinse  it  with  distilled  water,  blot  with  paper  towels  and 

immerse in the solution to be analysed, stir gently and wait until the stability of value, when the icon 

 

appears on the display, take the reading. 

 

4.4  Compensation of the barometric pressure 

Since the measurement of the partial pressure of dissolved oxygen is also related to barometric pressure, 

this  instrument,  thanks  to  the  integrated  barometric  sensor,  is  able  to  compensate  each  minimum 

variation. 

To  view  the  barometric  pressure measured  by  the  instrument,  press  the 

  key  during  measurement 

and scroll through the measurement parameters: %O

2

 

 mg/L 

 mbar. 

 

 

Содержание OXY Series

Страница 1: ...OXY 70 _ Ver 1 0 _ 02 2017 Serie OXY Portable oximeter OXY 70 Optical sensor O2 mg l Temp mbar The Future Starts Now...

Страница 2: ...calibration GLP function 6 4 Dissolved oxygen measurement 7 4 1 Before starting 7 4 2 Methods of measurement 7 4 3 Making the measurement 7 4 4 Compensation of the barometric pressure 7 4 5 Salinity c...

Страница 3: ...B connection 13 12 Calibration of temperature sensor 14 12 1 Useful information 14 12 2 Calibration procedure 14 13 Technical specifications 15 14 Accessories and spare parts 16 15 Disposal of electro...

Страница 4: ...sfer Model Description LDO 70 probe connection DIN multipin Connection for optical oxygen probe LDO70 2mt LDO70 10mt with built in temperature sensor USB port USB port for data transfer to the compute...

Страница 5: ...3 Switching off the instrument In the measurement mode press and hold for 3 seconds to turn off the instrument 1 3 4 Auto power off This meter powers off automatically after 20 minutes of inactivity...

Страница 6: ...mg l mbar Momentary press During measurement press to enter setup During calibration press to confirm when stable During setup press to select confirm the program Momentary press During measurement pr...

Страница 7: ...ven at low dissolved oxygen readings LDO70 sensor 2 1 Measuring principle On a membrane permeable to oxygen has been fixed a chemical substance called luminophore Inside the sensor a light source puls...

Страница 8: ...e The oximeter is calibrated and ready for use 3 2 Calibration with zero oxygen standard Normally it is sufficient to calibrate the instrument in the air at 100 as explained previously However in some...

Страница 9: ...ion GLP function On this instrument you can set an expiration of calibration after which the instrument displays an error message Er 7 and does not allow new measurements until a new calibration is pe...

Страница 10: ...solved O2 Concentration expressed in mg L equivalent of ppm mg L ppm During measurement press key to change measuring method 4 3 Making the measurement Remove the electrode from the protective cap rin...

Страница 11: ...robe LDO70 If the instrument does not calibrate or reading does not stabilize it is necessary to perform a maintenance of the probe in the following way 1 Check that the luminophore is clean in case r...

Страница 12: ...hore 6 The reading is out of the measuring limits Check the connection between the probe and the instrument Check the integrity of the probe Perform a new calibration Calibration expired Perform a new...

Страница 13: ...will be saved 8 1 Parameters of the setup menu and factory default settings Parameter Description Display Settings Factory default P5 1 Calibration at 0 0 P5 2 Salinity compensation 0 0 50 0 ppt 0 0 P...

Страница 14: ...rument is to remain unused for an extended period remove the batteries and leave it without Main Menu P5 1 Calibration at 0 P5 2 Salinity compensation P5 3 CAL Due P5 4 Last calibration data P5 5 Rest...

Страница 15: ...for automatic recording When you press the button short press the display will flash the icon and the instrument starts recording the values at the interval chosen To stop automatic recording press t...

Страница 16: ...and follow the wizard to install in order USB port Driver DataLink Software 70 The software DataLink 70 is available freely from www xsinstruments com 11 3USB connection Turn on the instrument connect...

Страница 17: ...atic bath pre thermostated together with the reference thermometer Wait 10 minutes for the stability if the reading is equal to the reference sensor 0 5 C then it is not necessary to perform any calib...

Страница 18: ...C 0 0 60 0 C Resolution 0 1 C 0 1 C Accuracy 0 5 C 0 5 C Automatic and manual temperature compensation Yes Yes only automatic Salinity measuring range 0 50 ppt 0 50 ppt Salinity compensation Yes manu...

Страница 19: ...O70 2MT LDO70 10MT 50010292 Standard zero 0 Oxygen single use 1 bottle for DO7 3MT LDO70 2MT LDO70 10MT For more accessories and spare parts visit the website www xsinstruments com 15 Disposal of elec...

Отзывы: