The Future Starts Now

 

OXY 70 

4.5  Salinity compensation (manual) 

The salinity of the sample to be measured influences the dissolved oxygen partial pressure. 

For  a  correct  measurement  you  must  set  the  salinity  value  of  the  sample.  If  you  make  oxygen 

measurements on salt water or sea water it is important to correct the measurement by setting the salinity 

value indicative of the sample. 

The  value  set  by  default  is  0  ppt,  to  change  access  to 

P5.2

  parameter  of  the  setup  menu  and  set  the 

desired value between 0 ... 50ppt. 

The average salinity of sea water is 35ppt. 

4.6  Automatic temperature Compensation ( ATC) 

The  measurement  of  the  partial  pressure  of  the  oxygen  is  strongly  influenced  by  temperature.  This 

instrument  allows  you  to  compensate  for  the  change  of  the  measurement  made  by  the  temperature 

thanks to a built in temperature sensor, the instrument will automatically recognize the presence and the 

display  shows  the  ATC  icon.  During  the  measurements  it  is  good  to  always  check  that  the  measured 

temperature is stable and aligned with the sample. 

 

5. Maintenance of the probe LDO70 

If the instrument does not calibrate or reading does not stabilize, it is necessary to perform a maintenance 

of the probe in the following way: 

 

1)  Check that the luminophore is clean, in case rinse externally with water. 

2)  The luminophore must be intact, undamaged, and without holes. 

3)  Unscrew the luminophore and check that the inside is dry, free from condensation and infiltration. 

If there is condensation or seepage, check the integrity of the O-ring seal of the luminophore and, 

if necessary, replace it. Dry thoroughly with a paper towel and screw the luminophore ensuring an 

airtight seal. 

 

After the maintenance make a new

 Calibration in air at 100%

. 

 

If even after the maintenance the probe does not calibrate then replace the luminophore with a new one. 

5.1  Replacing the luminophore 

The efficiency of the luminophore decreases with wear to the point that the instrument is no longer able to 

calibrate  itself,  in  which  case  its  replacement  is  necessary.  To  replace  the  luminophore  proceed  as 

follows: 

1)  Unscrew the luminophore. 

2)  Check the integrity of the inside of the probe. 

3)  Replace the O-ring with the new one supplied with new luminophore. 

4)  Screw the new luminophore ensuring an airtight seal. 

After the maintenance make a new 

Calibration of the oxygen sensor

 

 

Содержание OXY Series

Страница 1: ...OXY 70 _ Ver 1 0 _ 02 2017 Serie OXY Portable oximeter OXY 70 Optical sensor O2 mg l Temp mbar The Future Starts Now...

Страница 2: ...calibration GLP function 6 4 Dissolved oxygen measurement 7 4 1 Before starting 7 4 2 Methods of measurement 7 4 3 Making the measurement 7 4 4 Compensation of the barometric pressure 7 4 5 Salinity c...

Страница 3: ...B connection 13 12 Calibration of temperature sensor 14 12 1 Useful information 14 12 2 Calibration procedure 14 13 Technical specifications 15 14 Accessories and spare parts 16 15 Disposal of electro...

Страница 4: ...sfer Model Description LDO 70 probe connection DIN multipin Connection for optical oxygen probe LDO70 2mt LDO70 10mt with built in temperature sensor USB port USB port for data transfer to the compute...

Страница 5: ...3 Switching off the instrument In the measurement mode press and hold for 3 seconds to turn off the instrument 1 3 4 Auto power off This meter powers off automatically after 20 minutes of inactivity...

Страница 6: ...mg l mbar Momentary press During measurement press to enter setup During calibration press to confirm when stable During setup press to select confirm the program Momentary press During measurement pr...

Страница 7: ...ven at low dissolved oxygen readings LDO70 sensor 2 1 Measuring principle On a membrane permeable to oxygen has been fixed a chemical substance called luminophore Inside the sensor a light source puls...

Страница 8: ...e The oximeter is calibrated and ready for use 3 2 Calibration with zero oxygen standard Normally it is sufficient to calibrate the instrument in the air at 100 as explained previously However in some...

Страница 9: ...ion GLP function On this instrument you can set an expiration of calibration after which the instrument displays an error message Er 7 and does not allow new measurements until a new calibration is pe...

Страница 10: ...solved O2 Concentration expressed in mg L equivalent of ppm mg L ppm During measurement press key to change measuring method 4 3 Making the measurement Remove the electrode from the protective cap rin...

Страница 11: ...robe LDO70 If the instrument does not calibrate or reading does not stabilize it is necessary to perform a maintenance of the probe in the following way 1 Check that the luminophore is clean in case r...

Страница 12: ...hore 6 The reading is out of the measuring limits Check the connection between the probe and the instrument Check the integrity of the probe Perform a new calibration Calibration expired Perform a new...

Страница 13: ...will be saved 8 1 Parameters of the setup menu and factory default settings Parameter Description Display Settings Factory default P5 1 Calibration at 0 0 P5 2 Salinity compensation 0 0 50 0 ppt 0 0 P...

Страница 14: ...rument is to remain unused for an extended period remove the batteries and leave it without Main Menu P5 1 Calibration at 0 P5 2 Salinity compensation P5 3 CAL Due P5 4 Last calibration data P5 5 Rest...

Страница 15: ...for automatic recording When you press the button short press the display will flash the icon and the instrument starts recording the values at the interval chosen To stop automatic recording press t...

Страница 16: ...and follow the wizard to install in order USB port Driver DataLink Software 70 The software DataLink 70 is available freely from www xsinstruments com 11 3USB connection Turn on the instrument connect...

Страница 17: ...atic bath pre thermostated together with the reference thermometer Wait 10 minutes for the stability if the reading is equal to the reference sensor 0 5 C then it is not necessary to perform any calib...

Страница 18: ...C 0 0 60 0 C Resolution 0 1 C 0 1 C Accuracy 0 5 C 0 5 C Automatic and manual temperature compensation Yes Yes only automatic Salinity measuring range 0 50 ppt 0 50 ppt Salinity compensation Yes manu...

Страница 19: ...O70 2MT LDO70 10MT 50010292 Standard zero 0 Oxygen single use 1 bottle for DO7 3MT LDO70 2MT LDO70 10MT For more accessories and spare parts visit the website www xsinstruments com 15 Disposal of elec...

Отзывы: