6. Error messages
During calibration and measurement, the instrument displays error messages in case of problems:
Error messages
Error
Description
Solution
1
During calibration, the value is out of the
acceptable range
Calibrate at 0% with fresh standard
Check the integrity of the probe and
luminophore
Replace luminophore
2
During calibration appears when you
press the
button with unstable value
Wait for
icon
3
It appears if during the calibration the
value does not stabilize within 3 minutes
Check the integrity of the probe and
luminophore
Replace luminophore
6
The reading is out of the measuring limits
Check the connection between the probe and
the instrument
Check the integrity of the probe
Perform a new calibration
Calibration expired
Perform a new calibration in air
8
The LDO
sensor doesn’t communicate
with the instrument
Check the connection between the probe and
the instrument
Contact technical support
9
The built-in barometric sensor is not
working properly
Replace the batteries
Contact technical support
7. Restore to factory default setting
To restore the factory settings, access the parameter
P5.5
of setup menu.
Press
, the display will show
NO
flashing.
With the
or
keys choose
Yes
and confirm with the
key.
This feature can be very useful in case of errors or when making incorrect calibrations and / or have
calibration problems.