33
Rev. 1 24/01/2020
EN
yet.
•
Wait for the Conductivity value on the display to stabilize; when the icon
appears, use the keys
and
to adjust the value by entering that of the standard solution (ex: 5.00 µS/cm), as
indicated by the string “
ADJUST THE VALUE
” and by the icon
.
•
When the icon
reappears confirm the calibration point by pressing the key
.
•
The actual measured value flashes on the display and the updated cell constant is then shown.
•
The icon
appears, which indicates that the instrument is calibrated in the low conductivity range.
Automatically, the device returns to measure mode.
•
For each cell constant (P3.1) the instrument stores the calibration in order to allow the user, who uses
multiple sensors with different constants not to be forced to recalibrate each time.
The instrument recalls the last calibration with the respect to the selected parameters P3.1 (cell
constant) and P3.2 (type of calibration solutions).
Note:
If you are not aware of the exact compensation coefficient, to obtain an accurate calibration and
measurement set in P3.4
0.00 %/°C and then work by bringing the solutions exactly to the reference
temperature
Another method of working without temperature compensation is to use the appropriate thermal tables
shown on the most Conductivity solutions.
Important
: Always rinse the cell with distilled water before calibration and when switching from one
standard solution to another to avoid contamination.
Replace standard solutions frequently, especially low conductivity ones.
Contaminated or expired solutions can affect the accuracy and precision of the measurement.
•
Errors during calibration
•
NOT STABLE
:
The button
has been pressed with unstable signal. Wait for the icon
to
appear, to confirm the first point.
•
WRONG BUFFER
:
The buffer you are using is polluted or not part of the recognized families.
•
CALIBRATION TOO LONG
:
The calibration has exceeded the time limit, only the points calibrated up to
that moment will be stored.
•
Performing Conductivity measurement
•
Access the Setup menu for Conductivity to check the calibration, and if necessary, update the reading
parameters; press key
to return to measure mode.
•
Press
to scroll through the different screens of parameters until activating the Conductivity
parameter indicated by the icon
.
•
Connect the Conductivity cell to the BNC for Cond of the instrument (grey).