25
Rev. 1 24/01/2020
EN
recognize the chip, it works as a normal “analog” electrode.
•
Errors during calibration
•
NOT STABLE
:
The button
was pressed with still unstable signal. Wait for the icon
to appear
to confirm the point.
•
WRONG BUFFER
: The buffer is polluted or not part of the recognized families.
•
SLOPE OUT OF RANGE
: The slope of the sensor calibration line is out of the acceptable range 80 - 120%.
•
CALIBRATION TOO LONG
:
The calibration exceeded the time limit: only the points calibrated up to that
moment will be kept.
mV Parameter
pH 70 Vio; PC 70 Vio
•
In measure mode press the key
and move to the mV parameter indicated by the icon
.
•
The display shows the measurement in mV of the pH sensor.
•
The scrolling on the display of four red bands
means that the measurement is not stable yet.
•
Consider the measurement truthful only when the stability icon appears
.
Note
: This measurement is recommended to evaluate the sensor efficiency
ORP Parameter (Oxide-Reduction potential)
pH 70 Vio; PC 70 Vio
ORP sensors can be used on this series of devices to measure the Oxide-Reduction potential.
Connect the Redox electrode to the BNC type connector marked in green; if necessary, connect the
temperature probe to the RCA / CINCH Temp connector always marked with a green background.
It is possible to calibrate the sensor offset by performing automatic calibration on a predefined point. The
instrument automatically recognizes the
Redox solution 475 mV / 25 °C
; contact the local distributor to
proceed with the relevant purchase.
The instrument can correct the sensor offset by + 75 mV.
•
ORP parameter Setup
•
In measure mode press the key
to access the SETUP menu.
•
Use the directional keys to move to
ORP SETTINGS P2.0
and access the menu by pressing the key
.
•
Move with the keys
and
to select the program to access.
The table below shows the setup menu structure for the ORP parameter; for each program there are the
options that the user can choose and the default value: