21
Rev. 1 24/01/2020
EN
Access this menu to get information about the last calibration performed. The following screens will
automatically scroll on the display:
•
First screen: CALIBRATION DATE and TIME and beakers indicating the buffers used.
•
Second screen: OFFSET value of the electrode expressed in mV.
•
Third and possibly fourth screen: Slope% in the measuring range (one Slope% only if two calibration
points are performed, two Slope% if three points are performed).
Note
: The instrument accepts calibrations with pH electrodes with Slope% between 80 - 120% only.
Outside this range of acceptability, the instrument does not allow to end the calibration and displays the
error message
SLOPE OUT OF RANGE
P1.7 pH calibration deadline
Access this menu to set a calibration deadline; this option is very important in GLP protocols.
•
By default, no calibration deadline is set. Use the directional keys to select HOURS or DAYS and access
with button
. Use the directional keys to change the number that appears in the center of the
display, entering the hours or days that must elapse between two settings, and confirm with
.
•
If a calibration deadline is set, the icon
is present on the display in measure mode.
•
When the calibration deadline is activated, the instrument prevents further measurements.
The error icon
and the icon representing the calibration deadline
flash on the display.
The string “
MAKE A NEW CAL
” invites the user to perform a new calibration of the pH sensor to be
able to work again.
Press the key
to start the calibration.
P1.8 Reset of the pH parameter
If the instrument does not work perfectly or incorrect calibrations have been carried out, confirm
YES
with
the button
to take all the parameters of the pH menu back to the default settings.
IMPORTANT:
The factory reset of the parameters
DOES NOT
erase the stored data.
P1.9 Temperature calibration
All the instruments of this series are pre-calibrated for a correct temperature reading. However, if there is a
difference between the measured and the real temperature (usually due to a probe malfunction), it is
possible to perform an offset adjustment of + 5°C.
Use the keys
and
to correct the temperature offset value and confirm with
.
•
Automatic pH calibration
Example: three-point calibration with USA type buffers (7.00 / 4.01 / 10.01).
•
In
pH
measure mode
press the key
to enter calibration mode.
Содержание 70 Vio Series
Страница 1: ...1 Rev 1 24 01 2020 EN 70 Vio INSTRUCTIONS MANUAL pH Cond PC ...
Страница 2: ...2 Rev 1 24 01 2020 EN ...
Страница 5: ...5 Rev 1 24 01 2020 EN ...