10
Rev. 1 24/01/2020
EN
•
Datasheet
Series 70 Vio
pH
pH 70 Vio - PC 70 Vio
Measuring range
-2…16
Resolution / Accuracy
0.1, 0.01 / +0.02
Recognized calibration points and
buffers
AUTO:
1…3 / USA, NIST
CUS:
2 user values
Buffers indication
Yes
Calibration report
Yes
Automatic DHS recognition
Yes
Stability filter
Low – Med - High
mV
pH 70 Vio - PC 70 Vio
Range / Resolution
Range: -1000 … +1900 / Resolution: 0.1 / 1
ORP
pH 70 Vio - PC 70 Vio
Calibrations points
1 point / 475 mV
Conductivity
COND 70 Vio - PC 70 Vio
Range / Resolution
0,00 – 20,00 – 200,0 –
2000 μ
S /
2,00 – 20,00 – 200,0 mS
Automatic scale
Recognized calibration points and
buffers
1…5 / 84, 147,
1413 μ
S, 12.88, 111.8 mS,
1 user value
Reference temperature
15…30 °C
Temperature coefficient
0,00…10,00 %/°C
TDS
COND 70 Vio - PC 70 Vio
Measuring range / TDS Factor
0,1mg/l…200 gr/l / 0.40…1.00
Salinity
COND 70 Vio - PC 70 Vio
Measuring range
0,1ppm …100ppt
Temperature
pH 70 Vio - COND 70 Vio - PC 70 Vio
Measuring range
-10…110 °C
Resolution / Accuracy
0,1 / ± 0,5°C
Temperature compensation ATC
(NTC30K
Ω
) and MTC
0…100 °C
System
GLP with calibration Timer
Yes
Internal memory
1000 Dates
Display
High definition colours LCD
Brightness and contrast management
Manual and automatic with integrated sensor
IP protection
IP 57
Power supply
AA 1,5 V – 3 batteries / 5 V power supply with USB cable
Sound level during standard operation
< 80 dB
Environmental operating conditions
0 … +60 °C
Maximum permissible humidity
< 95 % non-condensing
Maximum altitude of use
2000 m
System dimensions
185 x 85 x 45 mm
System weight
400 g
Reference regulations
EMC 2014/30/UE
RoHS 2011/65/EU
EN 61326-1
EN 61010-1
Содержание 70 Vio Series