Setup for Conductivity parameter
•
In measure mode press button
to access the SETUP menu.
•
Use the directional keys to move to
COND SETTINGS P3.0
and access the menu by pressing the key
.
•
Move with keys
and
to select the program to access.
The table below shows the setup menu structure for the COND parameter; for each program, there are the
options that the user can choose and the default value:
Program
Description
Options
Factory Default
Settings
P3.1
CELL CONSTANT
0.1
- 1 - 10
1
P3.2
CALIBRATION METHOD
AUTOMATIC / CUSTOM
AUTOMATIC
P3.3
REFERENCE TEMPERATURE
15 … 30 °C
25 °C
P3.4
TEMP COMPENSATION FACTOR
0.00 … 10.00 %/°C
1.91 %/°C
P3.6
CALIBRATION DATA
-
-
P3.7
SET DUE CAL
NO – HOURS - DAYS
NO
P3.8
RESET SETTINGS
YES – NO
NO
P3.9
TEMPERATURE CAL
YES – NO
-
P3.1 Cell constant Selection
Choosing the right conductivity cell is a decisive factor for obtaining accurate and reproducible
measurements.
One of the most important parameters to consider is to use a sensor with the right cell constant in relation
to the solution under analysis.
The following table relates the sensor cell constant with the measurement range and the preferable
standard for calibration:
Access this setup menu to select the cell constant related to the sensor used:
•
0.1
•
1
- default-
•
10
•
For each of the 3 selectable cell constants, the instrument stores the calibrated points. By selecting the
cell constant, the previously performed calibration points are automatically recalled.