8.6. Track Manager Registers
The track manager registers are common for all cameras that share the same platform.
The track manager registers base address is: 0x5000.
Name
Offset
(hex)
R/W
Description
track.mode
0000:5000
RW
0 = off
1 = level
3 = level and gain
track.level
0000:5004
RW
Target tracking level
track.ref.x
0000:5008
RW
Reference frame used while tracking
track.ref.y
0000:500C
RW
track.ref.width
0000:5010
RW
track.ref.height
0000:5014
RW
Table 8-7 Track manager registers
8.7. XSP Registers
The XSP registers are common for all cameras that share the same platform. The XSP
registers base address is: 0x6000.
Name
Offset
(hex)
R/W
Description
XspVidPid
0000:6000
R
Internal ID
XspSerial
0000:6004
R
Serial nr
XspName
0000:6008
R
Camera name
XspVersion
0000:6028
R
XSP protocol version
XspFirmwareVersion
0000:602C
R
Internal version
Table 8-8 XSP registers
8.8. Bulk Memory Access Registers
Name
Offset
(hex)
R/W
Description
mem
0010:0000
RW
1MByte bulk memory
mem.size
0020:0000
R
Written data size
mem.crc32
0020:0004
R
Bulk memory CRC
Table 8-9 Bulk memory access registers