![Xenics Bobcat 320 Gated Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/xenics/bobcat-320-gated/bobcat-320-gated_user-manual_3256096032.webp)
User Manual
Bobcat 320 Gated
Company confidential.
This document is the property of Xenics. It may not be reproduced – completely or partially – or passed to a third
party without written permission from Xenics.
Xenics nv
T +32 16 38 99 00
Doc Ref: ENG-2012-UMN016
Ambachtenlaan 44
F +32 16 38 99 01
Issue: 002
BE-3001 Leuven
Belgium
www.xenics.com
Date: 09/11/2012
XF-104_02/20-01-2012
Page 32 of 34
8.3. Platform Registers
The platform registers are common for all cameras that share the same platform. The
system registers base address is: 0x2000.
Name
Offset
(hex)
R/W
Power- on
default
Description
mode
0000:2000
RW
15
See ‘mode flags’
setmode
0000:2004
RW
15
Set mode flag
clrmode
0000:2008
RW
15
Clear mode flag
roi.x
0000:2010
RW
0
Region of interest rectangle
roi.y
0000:2014
RW
0
roi.width
0000:2018
RW
384
roi.height
0000:201C
RW
288
roi.dir
0000:2020
RW
0
Bit 0 = horizontal flip
Bit 1 = vertical flip
level
0000:2024
RW
level offset
span
0000:2028
RW
gain value
trigger.mode
0000:2030
RW
0
See chap.
8.3.1 Trigger
trigger.delay
0000:2034
RW
0
trigger.skip
0000:2038
RW
0
Table 8-4 Platform registers
8.3.1.
Trigger
8.3.1.1. Trigger Mode
•
Bit 0 = Software trigger armed
•
Bit 1 = External trigger armed
•
Bit 2 = Fast rearm enabled (if disabled, multiple triggers in the same frame are
discarded)
•
Bit 3 = Edge(0)/Level(1)
•
Bit 4 = Rising, High(0) / Falling, Low(1).
8.3.1.2. Trigger Delay
The trigger delay is the amount of micro-seconds between the external trigger and the
actual trigger action.
Typically, there is an intrinsic jitter of ½ frame time.
8.3.1.3. Trigger Skip
The trigger skip is the amount of external triggers to be skipped.
i