Remote Operation
5-32
M370078-01
Figure 5-18 summarizes the Standard Event Status Register.
Figure 5-18
Summary of Standard Event Status Register
Table 5-10
Standard Event Status Register
Bit
Bit Weight
Bit Name
Description
0
1
Operation Complete
(OPC)
Set if
*OPC
command has been received and all
pending operations have been completed.
1
2
Request Control (RQC)
Not implemented. Always set to 0.
2
4
Query Error (QYE)
Not implemented.
3
8
Device Dependent Error
(DDE)
Set if there is a device-specific error.
See “Device-Specific Error List” on page B–5
for a list possible error codes and descriptions.
4
16
Execution Error (EXE)
Set if a program data element, following a header,
was evaluated by the power supply as outside of its
legal input range, or is otherwise inconsistent with
the power supply’s capabilities. Suggests that a
valid program message could not be properly
executed due to some power supply condition.
See “Execution Error List” on page B–4 for a list of
possible error codes and descriptions.
0
1
2
3
4
5
6
7
Standard Event
Status Register
(SESR)
+
SESR
Summary
Bit
...
0
1
2
3
4
5
6
7
Status Byte
Register
+
MSS
Summary
Bit
OPERation Status Summary Bit
QUEStionable Status Summary Bit
Error/Event Queue Status Flag
Not Used
Not Used
Not Used
Operation Complete
Not Used
Query Error
Device Dependent Error
Execution Error
Command Error
Not Used
Power On
Содержание Watt Series Programmable DC Power Supply XG 850
Страница 2: ......
Страница 3: ...XG 850 Watt Series Programmable DC Power Supply Operating Manual...
Страница 6: ......
Страница 18: ...xvi...
Страница 30: ...1 8...
Страница 52: ...Local Operation 3 8 M370078 01 Figure 3 3 Front Panel Menu System...
Страница 82: ...Local Operation 3 38 M370078 01 The SCPI command s for these instructions are SYSTem channel s REMote STATe...
Страница 86: ...Local Operation 3 42 M370078 01 Figure 3 5 Split Supply Operation V V Common Power Supply Power Supply Load...
Страница 126: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370078 01...
Страница 198: ...5 72...
Страница 262: ...A 36...
Страница 284: ...IX 4...
Страница 285: ......