Specification
OUTPUT
Voltage Range:
0V to 18V minimum
Current Range:
0A to 10A minimum
Output Voltage Setting:
By coarse and fine controls.
Output Current Setting:
By single logarithmic control.
Operating Mode:
Constant voltage or constant current with automatic cross-over.
Output Switch:
Electronic. Preset voltage and current displayed when off.
Output Terminals:
4mm terminals on 19mm (0.75”) pitch.
Output Impedance:
Typically <5m
Ω
in constant voltage mode.
Typically >10k
Ω
in constant current mode.
Output Protection:
Output will withstand up to 22V forward voltage.
Reverse protection by diode clamp for reverse currents up to 3A.
Load & Line Regulation:
<0.01% of maximum output for a 90% load change or 10% line
change.
Ripple & Noise
(20MHz bandwidth):
Typically <2mVrms, <10mV pk-pk, constant voltage mode.
Transient Response:
<200µs to within 50mV of set level for 90% load change.
Temperature Coefficient:
Typically <100ppm/°C
Status Indication:
Output on lamp. Constant current mode lamp.
METER SPECIFICATIONS
Meter Types:
Dual 4 digit meters with 14mm (0.56") LEDs. Reading rate 4 Hz.
Meter Resolutions:
10mV, 10mA
Meter Accuracies:
Voltage 0.3% of reading ± 2 digits
Current 0.6% of reading ± 2 digits
GENERAL
AC Input:
Factory set for: 110V - 120V AC ±10% or 220V - 240V AC ±10%,
50/60Hz. Installation Category II.
Power Consumption:
400VA max.
Operating Range:
+5ºC to +40ºC, 20% to 80% RH.
Storage Range:
−
40ºC to + 70ºC.
Environmental:
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Safety:
Complies with EN61010
−
1.
EMC:
Complies with EN61326.
Size:
140 x 160 x 320mm (WxHxD).
Weight: 2.8kg
3