Operator’s Manual
FEATURES:
Control Box Connections:
The control box disperses the power from the battery to the
leg, lift arm, and handset functions.
Emergency Off:
In case of emergency, push the large red button to cut
power to unit.
Ensure red button is disengaged in order to return unit to
functionality. To disengage, turn the red button a half turn
clockwise. This will re-activate the spring-loaded button.
Manual Emergency Down:
In the event there is not enough power to lower the patient,
pull up on the red handle at the neck of the actuator as
shown. The lift arm will slowly lower the patient
(Must have
weight to lower).
Powered Emergency Down:
In the event the handset does not function, push the button
on the control box to lower the lift arm.
Leg Actuators
Lift Arm Actuator
Mains
Cable
Handset
Figure 3