PurCon
Maintenance
73
On the following pages, cleaning the PurCon with hygienic cleaner and
with detergent concentrate are described in detail.
Method 1:
Cleaning using hygienic
cleaner
(e. g. Dan Klorix)
To clean the PurCon, carry out the following steps:
Fig. 14
Both valves closed
8
Disable the feed transport.
9
Switch off the permeate pump (selector switch to position 0).
10
Switch off the power switch of the PurCon.
11
Close the manual Inlet valve by turning it.
12
By opening the Outlet valve, empty the sample preparation
instrument.
13
After approx. 30 s, close the Outlet valve again.