Maintenance
PurCon
72
5.3
Cleaning the PurCon
5.3.1
Housing
Wipe the PurCon with a damp cloth.
If necessary, neutral soap can be used for cleaning.
Caution
Do not clean the PurCon with water jet or steam jet.
5.3.2
Sample preparation
Caution
Only qualified specialist personnel are allowed to perform cleaning
activities.
If the analyzers do not obtain sufficient permeate, they generate a cor-
responding error message. In this case, the inside of the PurCon has
to be cleaned. It is cleaned chemically by the effect of 1 % sodium
hypochlorite solution. The cleaning solution can be produced from var-
ious detergent concentrates by dilution. A selection of possible deter-
gent concentrates are given in the chapter 8 A
CCESSORIES
AND
CONSUMABLE
MATERIALS
.
Wear protective goggles!
Wear protective gloves!
Wear protective clothing!
Содержание PurCon
Страница 1: ...PurCon Sample preparation ba43107e10 11 2004 47 Operating manual...
Страница 4: ...Contents PurCon 50...
Страница 12: ...Safety instructions PurCon 58...
Страница 22: ...Operation PurCon 68...
Страница 34: ...Maintenance PurCon 80...
Страница 36: ...Maintenance PurCon 82...