pH value
pH 7310
28
ba75929e02
10/2011
5.2.4 Manual calibration (ConCal)
Make sure that in the sensor menu,
Buffer
menu, the
ConCal
buffer set is
selected (see section 7.1.1 C
page 38).
Use any one to five buffer solutions in ascending or descending order.
21. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
22. If necessary, use
<M>
to finish calibration or
Continue calibrating using the next buffer with
<ENTER>
.
Calibration is automatically completed after the last buffer of a buf-
fer set has been measured. Then the calibration record is dis-
played.
The calibration line is determined by linear regression.
If single-point calibration was set in the menu, the calibration proce-
dure is automatically finished with the measurement of buffer solu-
tion 1 and the calibration record is displayed.
1.
Connect the pH electrode to the meter.
The pH measuring window is displayed.
2.
Keep the buffer solutions ready.
When measuring without temperature sensor:
Temper the buffer solutions or measure the current temperature.
3.
In the measured value display, select the measured parameter pH or
mV with
<M>
.
4.
Start the calibration with
<CAL>
.
The calibration display appears.
5.
Thoroughly rinse the electrode with deionized water.
pH
0.0
24.8 °C
ConCal
1
01.07.2011
08:00
Buffer
mV
Содержание pH 7310
Страница 1: ...ba75929e02 10 2011 pH 7310 pH 7310P Operating manual pH meter pH 7310P pH 7310 Printermodul...
Страница 66: ...Firmware update pH 7310 66 ba75929e02 10 2011...
Страница 70: ...Index pH 7310 70 ba75929e02 10 2011...
Страница 71: ......