pH 7310
pH value
ba75929e02
10/2011
23
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
1.
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
The HOLD function is active.
You can terminate the
Stability control
function and the HOLD func-
tion with
<AR>
or
<M>
at any time.
2.
Using
<ENTER>
, activate the
Stability control
function manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable mea-
sured value is recognized. The progress bar disappears and the display
of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the
Stability control
function manu-
ally with
<ENTER>
at any time. If the
Stability control
function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3.
Using
<ENTER>
, start a further measurement with stability control.
or
Release the frozen measured value again with
<AR>
or
<M>
.
The [AR] status display disappears. The display switches back to the
previous indication.
Measured
parameter
Time interval
Stability in the time interval
pH value
15 seconds
Δ
pH: Better than 0.01
Temperature
15 seconds
Δ
T (° C): Better than 0.02
Содержание pH 7310
Страница 1: ...ba75929e02 10 2011 pH 7310 pH 7310P Operating manual pH meter pH 7310P pH 7310 Printermodul...
Страница 66: ...Firmware update pH 7310 66 ba75929e02 10 2011...
Страница 70: ...Index pH 7310 70 ba75929e02 10 2011...
Страница 71: ......