![wtw pH 3300i Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/wtw/ph-3300i/ph-3300i_operating-manual_1000653044.webp)
Operation
pH 3300i/3400i
44
ba75682e02
03/2008
4.5.2
Recorder (analog output, pH 3400i)
You can transmit data to a recorder via the analog output. Connect the ana-
log output to the recorder via the AK323 interface cable.
The data output automatically switches to
Recorder output.
Socket
assignment
Note
The analog output is activated automatically in the cable by connecting 2 and
3.
At the analog output, there is a low-impedance electrode signal (correspond-
ing to 50 ... 62 mV/pH).
1 free
2 Plug coding
3 Ground
4 Analog output
1
2
3
4
RS 232
REC
(internal resistance < 5 Ohm)
Содержание pH 3300i