Operation
pH 3300i/3400i
30
ba75682e02
03/2008
Nernst slope (-59.2 mV/pH at 25 °C) and the specified asymmetry of the elec-
trode.
8
To continue the
two-point calibration
, thoroughly rinse the elec-
trode with deionized water.
9
If necessary, enter the temperature of the second buffer solution with
<
▲
> <
▼
>
.
10
Immerse the pH electrode in the second buffer solution.
11
Press the
<RUN/ENTER>
key. The second pH value appears on the
display.
12
Set the measured value to the nominal pH value of the buffer solution
(at the current temperature) with
<
▲
> <
▼
>
.
13
When the measured value is stable, press the
<RUN/ENTER>
key.
The value of the slope (mV/pH) appears on the display. The probe
symbol shows the evaluation of the current calibration.
14
Press the
<RUN/ENTER>
key. The value of the asymmetry (mV) ap-
pears on the display.
AutoCal DIN
AutoCal TEC
6
mV/pH
mol/L
8
Auto Store
RCL
LoBat
CalError
ARng AR
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
O
SL
°C
TP
pH
ConCal
Содержание pH 3300i
Страница 4: ...List of contents pH 3300i 3400i 4 ba75682e02 03 2008 7 Technical data 57 8 Lists 61...
Страница 31: ...pH 3300i 3400i Operation 31 ba75682e02 03 2008 15 To return to the measuring mode Press the M key...
Страница 56: ...What to do if pH 3300i 3400i 56 ba75682e02 03 2008...
Страница 60: ...Technical data pH 3300i 3400i 60 ba75682e02 03 2008...
Страница 66: ...Lists pH 3300i 3400i 66 ba75682e02 03 2008...
Страница 68: ...Index pH 3300i 3400i 68 ba75682e02 03 2008...