Operation
pH 3300i/3400i
28
ba75682e02
03/2008
Three-point
calibration
Note
You can prematurely terminate the three-point calibration by pressing
<M>
.
The values of the two-point calibration for slope and asymmetry will then re-
main stored.
12
Press the
<RUN/ENTER>
key
to continue the
three-point calibration
.
Cd3
appears on the display.
13
Thoroughly rinse the electrode with distilled water.
14
If necessary, enter the temperature of the third buffer solution with
<
▲
> <
▼
>
.
15
Immerse the pH electrode in the third buffer solution.
16
Press the
<RUN/ENTER>
key.
The
AR
display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on
the display.
When the measured value is stable,
AR
disappears.
The instrument displays the value of the slope (mV/pH) and the eval-
uation of the calibration in calibration range II.
17
Press the
<RUN/ENTER>
key.
The instrument displays the value of the asymmetry in calibration
range II. The three-point calibration is completed.
18
To return to the measuring mode: Press the
<M>
key.
Содержание pH 3300i
Страница 4: ...List of contents pH 3300i 3400i 4 ba75682e02 03 2008 7 Technical data 57 8 Lists 61...
Страница 31: ...pH 3300i 3400i Operation 31 ba75682e02 03 2008 15 To return to the measuring mode Press the M key...
Страница 56: ...What to do if pH 3300i 3400i 56 ba75682e02 03 2008...
Страница 60: ...Technical data pH 3300i 3400i 60 ba75682e02 03 2008...
Страница 66: ...Lists pH 3300i 3400i 66 ba75682e02 03 2008...
Страница 68: ...Index pH 3300i 3400i 68 ba75682e02 03 2008...