![wtw pH 3300i Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/wtw/ph-3300i/ph-3300i_operating-manual_1000653016.webp)
Operation
pH 3300i/3400i
16
ba75682e02
03/2008
AutoRead
(drift control)
The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of the
measured values.
For identical measurement conditions, the following criteria apply:
Note
The current AutoRead measurement can be terminated at any time (accept-
ing the current value) by pressing <RUN/ENTER>.
Reproducibility
Response time
Better than 0.02
> 30 seconds
1
Call up the pH measuring mode with
<M>
and
<
▲
> <
▼
>
.
2
Activate the AutoRead function with
<AR>
.
The current measured value is frozen (hold function).
3
Start AutoRead with
<RUN/ENTER>
.
AR
flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4
If necessary, start the next AutoRead measurement with
<RUN/EN-
TER>
.
5
To terminate the AutoRead function: Press the
<AR>
key.