pH 3300i/3400i
Operation
19
ba75682e02
03/2008
). In order to view the calibration data, switch the measuring instru-
ment off and switch it on again.
Printing the
calibration
record (3400i)
The calibration protocol contains the calibration data of the current calibration.
You can transmit the calibration protocol to a printer via the serial interface
(see page 40).
Note
You can automatically print a calibration protocol after the calibration. To do
so, connect a printer to the interface according to section 4.5.3 before cali-
brating. After a valid calibration, the record is printed.
Sample printout:
CALIBRATION PROTOCOL
02.03.99 14:19
Device No.: 12345678
Calibration pH
Cal time: 01.03.99 / 15:20
Cal interval: 7d
AutoCal DIN Tauto
Buffer 1 1.679
Buffer 2 4.008
Buffer 3 6.865
Buffer 4 9.180 *
C1 174.1 mV 25.0°C
C2 -133.3 mV 25.0°C
S1 -59.4 mV/pH
ASY1 - 4 mV
Probe: +++