![wtw Multi 350i Скачать руководство пользователя страница 73](http://html1.mh-extra.com/html/wtw/multi-350i/multi-350i_operating-manual_1000622073.webp)
Multi 350i
Operation
71
Settings
Menu item
Setting/
function
Description
Data storage
/
Manual data storage
/
Display
-
Displays in pages all mea-
suring datasets that corre-
spond to the filter settings.
Further options:
!
Scroll through the
datasets with
<
▲
> <
▼
>
.
!
Output the displayed
dataset to the interface
with
<PRT>
.
!
Quit the display with
<ESC>
.
Data storage
/
Manual data storage
/
Output to RS232
-
Outputs to the interface all
measuring datasets that
correspond to the filter set-
tings. The output takes
place in ascending order of
the ID number.
The process can take sev-
eral minutes. To terminate
the process prematurely,
press
<ESC>
.
Data storage
/
Manual data storage
/
Data filter
-> see expla-
nations below
this table
Allows to set certain filter
criteria in order to display
and output them to the in-
terface datasets.
Data storage
/
Manual data storage
/
Erase
-
Erases the entire contents
of the selected measuring
data storage, independent
of the filter settings.
Note:
All calibration data remains
stored when performing
this action.
Содержание Multi 350i