Operation
Multi 350i
64
Calibration evaluation
After the calibration, the measuring instrument automatically evaluates
the current status of the calibration. The evaluation appears on the dis-
play and in the calibration record.
Determining the
cell constant
For this calibration procedure, the
Measuring cell
setting must be set to
cal
in the measuring menu. Proceed as follows to determine the cell
constant:
Display
Calibration
record
Cell constant [cm
-1
]
+++
within the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 1.200 cm
-1
Error
Error
outside the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 1.200 cm
-1
Perform error elimination ac-
cording to chapter 6 W
1
Connect a conductivity measuring cell to the measuring instru-
ment.
2
In the measured value display, select the conductivity measur-
ing window with
<
▲
> <
▼
>
and
<M>
.
3
Start the calibration with
<CAL>
.
The calibration display appears.
4
Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
i
Cell constant = 0.475 1/cm
i
Immerse sensor in solution
Continue
<CAL>