![wtw Multi 350i Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/wtw/multi-350i/multi-350i_operating-manual_1000622029.webp)
Multi 350i
Operation
27
4.4.2
Measuring the pH value
AutoRead
(Drift control)
The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured vari-
able flashes until a stable measured value is available.
Criteria
With identical measurement conditions, the following applies:
1
Perform the preparatory activities according to section 4.6.1.
2
Immerse the pH electrode in the test sample.
3
Select the pH or mV display with
<M>
.
pH
6.949
24.8
°C
Measured
variable
Reproducibility
Response time
pH value
Better than 0.01
> 30 seconds