Multi 350i
Operation
47
Continuing for two-point
calibration
11
Wait for the end of the AutoRead measurement or accept the
calibration value with
<MENU/ENTER>
.
The calibration display for the next standard solution appears.
i
Standard = 0.010 mg/l
i
U = 0.5 mV
Terminate AutoRead
ISE <CAL>
i
#1 0.010 mg/l 24 °C
i
Immerse sensor in std. 2
Set temperature:
24 °C
Continue
Set standard:
0.020 mg/l
ISE <CAL>
12
Thoroughly rinse the electrode with distilled water.
13
Immerse the electrode in standard solution 2.
14
Select the
Set standard
setting with
<
▲
> <
▼
>
and press
<MENU/ENTER>
.
15
Set the concentration of the standard solution with
<
▲
> <
▼
>
and press
<MENU/ENTER>
.
16
Measure the temperature of the standard solution using a ther-
mometer.
17
Select the
Set temperature
setting with
<
▲
> <
▼
>
and press
<MENU/ENTER>
.
18
Set the temperature with
<
▲
> <
▼
>
and press
<MENU/ENTER>
.
19
Select
Continue
with
<
▲
> <
▼
>
and press
<MENU/ENTER>
.
The standard solution is measured.
The measured value is checked for stability (AutoRead).