Technical data
Cond 730
68
No compensation:
Accuracy
± 0.5 %
Non-linear compensation
:
Accuracy
Test sample temperature
± 0.5 %
0 °C ... 35 °C
according to EN 27 888;
± 0.5 %
35 °C ... 50 °C
extended nLF function according to
WTW measurements
Linear compensation
:
Accuracy
Test sample temperature
± 0.5 %
10 °C ... 75 °C
(the accuracy percentage always refers
to the measured value!)
SAL
Range 0.0 ... 42.0
Accuracy
Test sample temperature
± 0.1
5 °C ... 25 °C
± 0.2
25 °C ... 30 °C
TDS [mg/l]
1
T [°C]
NTC 30:
Accuracy
± 0.1
PT 1000:
Accuracy
Operating temperature
± 0.5
0 °C ... 15 °C
± 0.1
15 °C ... 35 °C
± 1
35 °C ... 55 °C
T [°F]
NTC 30:
Accuracy
± 0.2
PT 1000:
Accuracy
Operating temperature
± 0.9
32 °F ... 59 °F
± 0.2
59 °F ... 95 °F
± 1.8
95 °F ... 131 °F
Accuracy
(± 1 digit)