![wtw inoLab Cond 730 Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/wtw/inolab-cond-730/inolab-cond-730_operating-manual_1000595050.webp)
Operation
Cond 730
50
4.6.3 PC/external printer (RS232 interface)
You can transmit data to a PC or an external printer via the
RS232 interface.
Connect the interface to the instrument via the AK340/B
cable (PC) or AK325/S cable (external printer).
The data output switches automatically to RS232.
Note
The RS232 interface is not galvanically isolated.
If it is connected to an earthed PC/printer, measurements
cannot be made in earthed media as this would give incor-
rect results!
Set up the following transmission data on the PC/printer:
Baud rate
Selectable between:
1200, 2400, 4800, 9600
Handshake
RTS/CTS + Xon/Xoff
PC only:
Parity
None
Data bits
8
Stop bits
1
Socket assignment
1 CTS
2 RxD
3 Ground
4 TxD
1
2
3
4
RS 232
REC
Содержание inoLab Cond 730
Страница 1: ......
Страница 8: ...Overview Cond 730 8...
Страница 12: ...Safety Cond 730 12...
Страница 16: ...Commissioning Cond 730 16...
Страница 66: ...What to do if Cond 730 66...
Страница 78: ...Lists Cond 730 78...